User's Manual
Power Delta Calibration
68P64115A18–1
Mar 2003
1X SC 4812T Lite BTS Optimization/ATP Software Release R2.16.1.x
DRAFT
H-8
Table H-2: Advantest R3267 Power Delta Calibration Procedure
Step Action
41
Calculate the Power Calibration Delta value. The delta value is the power meter measurement minus
the Advantest measurement.
Delta = A – B
Example: Delta = –0.7 dBm – (–1.25 dBm) = 0.55 dB
Example: Delta = 0.26 dBm – 0.55 dBm = –0.29 dBm
These examples are included to show the mathematics and do not represent actual readings.
NOTE
Add this delta value to the TX Cable Loss value during In–Service Calibration (see step 4 in
Table H-6).
Figure H-3: Delta Calibration Setup – Advantest R3562 to HP437
Power
Sensor
Advantest R3562 and R3267
Short RF Cable
HP437B
SENSOR
RF OUT
Figure H-4: Delta Calibration Setup –
Advantest R3562 to HP437
Figure H-5: Delta Calibration Setup – Advantest R3562 to R3267
Advantest R3562 and R3267
Short RF Cable
RF OUT
RF IN
H