Troubleshooting: Installation continued
Jan 2002
6-3
SC4812ET BTS Optimization/ATP CDMA LMF
Cannot Communicate to
Communications Analyzer
Follow the procedure in Table 6-3 to troubleshoot a communication
analyzer failure.
Table 6-3: Troubleshooting a Communications Analyzer Communication Failure
Step Action
1 Verify analyzer is connected to LMF with GPIB adapter.
2 Verify cable setup.
3 Verify the GPIB address is set to 18.
4 Verify the GPIB adapter DIP switch settings are correct. Refer to Test Equipment setup section
for details.
5 Verify the GPIB adapter is not locked up. Under normal conditions, only 2 green LEDs must be
ON (Power and Ready). If any other LED is continuously ON, then power-cycle the GPIB
Box and retry.
6 If a Hyperterm window is open for MMI, close it.
7 Verify the LMF GPIB address is set to 18
8 Verify the analyzer is in Talk and Listen not Control mode.
6