User Manual

Troubleshooting68P09255A57-2
Aug 2002
SC4812ET Optimization/ATP Manual Software Release R16.1.x.x
PRELIMINARY
6-5
Table 6-6: Troubleshooting Device Enable (INS) Failure
ActionStep
3 Ensure primary CSM is in INS_ACT state.
NOTE
MCCs will not go INS without the CSM being INS.
4 Verify 19.6608 MHz CSM clock; MCCs will not go INS otherwise.
5 The BBX should not be enabled for ATP tests.
6 If MCCs give invalid or no system time, verify the CSM is enabled.
LPA Errors
Follow the procedure in Table 6-7 to troubleshoot any LPA errors.
Table 6-7: LPA Errors
Step Action
1 If LPAs continue to give alarms, even after cycling power at the circuit breakers, then connect an
MMI cable to the LPA and set up a Hyperterminal connection. Enter
ALARMS in the
Hyperterminal window. The resulting LMF display may provide an indication of the problem.
(Call Field Support for further assistance.)
Bay Level Offset Calibration Failure
Follow the procedure in Table 6-8 to troubleshoot a BLO calibration
failure.
Table 6-8: Troubleshooting BLO Calibration Failure
Step Action
1 Verify the Power Meter is configured correctly (see the test equipment setup section) and
connection is made to the proper TX port.
2 Verify the parameters in the bts-#.cdf file are set correctly for the following bands:
For 1900 MHz:
BandClass=1; FreqBand=16
For 800 MHz:
BandClass=0; FreqBand=8
3 Verify that no LPA in the sector is in alarm state (flashing red LED). Reset the LPA by pulling
the circuit breaker, and after 5 seconds, pushing back in.
4 Re-calibrate the Power Meter and verify it is calibrated correctly with cal factors from sensor
head.
5 Verify GPIB adapter is not locked up. Under normal conditions, only 2 green LEDs must be
ON (Power and Ready). If any other LED is continuously ON, power-cycle (turn power off
and on) the GPIB Box and retry.
6 Verify sensor head is functioning properly by checking it with the 1 mW (0 dBm) Power Ref
signal.
. . . continued on next page
6