User's Manual
Setting GPIB Addresses68P09255A69-3
Aug 2002
1X SC4812ET Lite BTS Optimization/ATP Software Release 2.16.1.x
PRELIMINARY
F-3
Setting GPIB Addresses
Procedures for Verifying and Setting CDMA Support Equipment GPIB
Addresses
The following procedures cover verification and changing GPIB
addresses for the various items of CDMA test equipment supported by
the LMF.
Agilent E4406A Transmitter Tester GPIB Address
Figure F-1: Setting Agilent E4406A GPIB Address
System
Key
Bk Sp
Key
Enter
Key
Data Entry
Keypad
Softkey
Buttons
Softkey Label
Display Area
Active Function
Area
Refer to Figure F-1 and follow the procedure in Table F-1 to verify and,
if necessary, change the Agilent E4406A GPIB address.
Table F-1: Verify and Change Agilent E4406A GPIB Address
Step Action
1 In the SYSTEM section of the instrument front panel, press the System key.
- The softkey labels displayed on the right side of the instrument screen will change.
2 Press the Config I/O softkey button to the right of the instrument screen.
- The softkey labels will change.
- The current instrument GPIB address will be displayed below the GPIB Address softkey label.
3 If the current GPIB address is not set to 18, perform the following to change it:
3a - Press the GPIB Address softkey button.
-- In the on-screen Active Function Area, GPIB Address will be displayed followed by the
current GPIB address.
. . . continued on next page
F