User's Manual
Table Of Contents
- Table of Contents
- List of Figures
- List of Tables
- Foreword
- FCC Requirements
- General Safety
- Revision History
- Patent Notification
- Chapter 1 Introduction
- Chapter 2 Preliminary Operations
- Chapter 3 Optimization/Calibration
- Optimization/Calibration - Introduction
- Isolate Span Lines/Connect LMF
- Preparing the LMF
- Using CDMA LMF
- Download the BTS
- CSM System Time/GPS and LFR/HSO Verification
- Test Equipment Set-up
- Test Set Calibration
- Bay Level Offset Calibration
- RFDS Set-up and Calibration
- BTS Redundancy/Alarm Testing
- BTS Alarms Testing
- Preliminary Operations: Overview
- Pre-Power-up Tests
- Initial Power-up Tests
- Chapter 4 Automated Acceptance Test Procedure
- Automated Acceptance Test Procedures - All-inclusive TX & RX
- TX Output Acceptance Tests: Introduction
- TX Spectral Purity Transmit Mask Acceptance Test
- TX Waveform Quality (rho) Acceptance Test
- TX Pilot Time Offset Acceptance Test
- TX Code Domain Power/Noise Floor Acceptance Test
- RX Frame Error Rate (FER) Acceptance Test
- Generate an ATP Report
- Chapter 5 Prepare to Leave the Site
- Chapter 6 Basic Troubleshooting
- Basic Troubleshooting Overview
- Troubleshooting: Installation
- Troubleshooting: Download
- Troubleshooting: Calibration
- Troubleshooting: Transmit ATP
- Troubleshooting: Receive ATP
- Troubleshooting: CSM Check-list
- C-CCP Backplane Troubleshooting
- Module Front Panel LED Indicators and Connectors
- Basic Troubleshooting - Span Control Link
- Appendix A Data Sheets
- Appendix B PN Offset/I & Q Offset Register Programming Information
- Appendix C FRU Optimization/ATP Test Matrix
- Appendix D BBX Gain Set Point vs. BTS Output
- Appendix E CDMA Operating Frequency
- Appendix F PCS Interface Setup for Manual Testing
- Appendix G VSWR
- Appendix H Download ROM Code
- Appendix I In-Service Calibration
- Appendix J GPIB Addresses
- Introduction
- HP437 Power Meter GPIB Address
- Gigatronics 8541C Power Meter GPIB Address
- Motorola CyberTest GPIB Address
- HP8935 Test Set GPIB Address
- Setting HP8921A and HP83236A/B GPIB Address
- Advantest R3465 GPIB Address
- RS232 GPIB Interface Box
- Advantest R3267 Spectrum Analyzer GPIB Address
- Agilent E4406A Transmitter Tester GPIB Address
- Index
RFDS Set-up and Calibration
68P09255A61-4
Aug 2002
1X SC 4812T BTS Optimization/ATP Software Release 2.16.1.x
FIELD TRIAL
3-84
Table 3-41: RFDS Parameter Settings
Step Action
NOTE
Log out of the BTS prior to performing this procedure.
1 Using a text editor, verify the following fields are set correctly in the bts-#.cdf file
(1 = GLI based RFDS; 2 = Cobra RFDS).
EXAMPLE:
Asu1Equip = 1
Asu2Equip = 0 (1 if system is non-duplexed)
Mc1Equip = 0
Mc2Equip = 0
Mc3Equip = 0
Mc4Equip = 0
RfdsEquip = 2
TestOrigDN = ’123456789’
TsuEquip = 1
NOTE
The above is an example of the bts-#.cdf file that should have been generated by the OMC and
copied to the LMF. These fields will have been set by the OMC if the RFDSPARM database is
modified for the RFDS.
2 Save and/or quit the editor. If any changes were made to these fields, data will need to be downloaded
to the GLI (see Step 3, otherwise proceed to Step 4).
3 To download to the GLI, click on the Device menu and select the Download>Code/Data menu item
(selected devices do not change color when data is downloaded).
A status report window displays the status of the download.
Click OK to close the status report window.
! CAUTION
After downloading data to the GLI, the RFDS LED slowly begins flashing red and green for
approximately 2-3 minutes. DO NOT attempt to perform any functions with the RFDS until the LED
remains green.
4 Status the RFDS TSU.
A status report window displays the software version number for the TSIC and SUA.
NOTE
If the LMF yields an error message, check the following:
Ensure the AMR cable is correctly connected from the BTS to the RFDS.
Verify the RFDS has power.
Verify the RFDS status LED is green.
Verify fields in the bts-#.cdf file are correct (see Step 1).
Status the MGLI and ensure the device is communicating (via Ethernet) with the LMF, and the
device is in the proper state (INS).
3