User's Manual
Table Of Contents
- Table of Contents
- List of Figures
- List of Tables
- Foreword
- FCC Requirements
- General Safety
- Revision History
- Patent Notification
- Chapter 1 Introduction
- Chapter 2 Preliminary Operations
- Chapter 3 Optimization/Calibration
- Optimization/Calibration - Introduction
- Isolate Span Lines/Connect LMF
- Preparing the LMF
- Using CDMA LMF
- Download the BTS
- CSM System Time/GPS and LFR/HSO Verification
- Test Equipment Set-up
- Test Set Calibration
- Bay Level Offset Calibration
- RFDS Set-up and Calibration
- BTS Redundancy/Alarm Testing
- BTS Alarms Testing
- Preliminary Operations: Overview
- Pre-Power-up Tests
- Initial Power-up Tests
- Chapter 4 Automated Acceptance Test Procedure
- Automated Acceptance Test Procedures - All-inclusive TX & RX
- TX Output Acceptance Tests: Introduction
- TX Spectral Purity Transmit Mask Acceptance Test
- TX Waveform Quality (rho) Acceptance Test
- TX Pilot Time Offset Acceptance Test
- TX Code Domain Power/Noise Floor Acceptance Test
- RX Frame Error Rate (FER) Acceptance Test
- Generate an ATP Report
- Chapter 5 Prepare to Leave the Site
- Chapter 6 Basic Troubleshooting
- Basic Troubleshooting Overview
- Troubleshooting: Installation
- Troubleshooting: Download
- Troubleshooting: Calibration
- Troubleshooting: Transmit ATP
- Troubleshooting: Receive ATP
- Troubleshooting: CSM Check-list
- C-CCP Backplane Troubleshooting
- Module Front Panel LED Indicators and Connectors
- Basic Troubleshooting - Span Control Link
- Appendix A Data Sheets
- Appendix B PN Offset/I & Q Offset Register Programming Information
- Appendix C FRU Optimization/ATP Test Matrix
- Appendix D BBX Gain Set Point vs. BTS Output
- Appendix E CDMA Operating Frequency
- Appendix F PCS Interface Setup for Manual Testing
- Appendix G VSWR
- Appendix H Download ROM Code
- Appendix I In-Service Calibration
- Appendix J GPIB Addresses
- Introduction
- HP437 Power Meter GPIB Address
- Gigatronics 8541C Power Meter GPIB Address
- Motorola CyberTest GPIB Address
- HP8935 Test Set GPIB Address
- Setting HP8921A and HP83236A/B GPIB Address
- Advantest R3465 GPIB Address
- RS232 GPIB Interface Box
- Advantest R3267 Spectrum Analyzer GPIB Address
- Agilent E4406A Transmitter Tester GPIB Address
- Index
Test Set Calibration68P09255A61-4
Aug 2002
1X SC 4812T BTS Optimization/ATP Software Release 2.16.1.x
FIELD TRIAL
3-63
Table 3-28: Cable Calibration
Step Action
1 From the Util menu, select Cable Calibration.
A Cable Calibration window is displayed.
2 Enter a channel number(s) in the Channels box.
NOTE
Multiple channels numbers must be separated with a comma, no space (i.e., 200,800). When two
or more channels numbers are entered, the cables are calibrated for each channel. Interpolation is
accomplished for other channels as required for TX calibration.
3 Select TX and RX Cable Cal, TX Cable Cal, or RX Cable Cal in the Cable Calibration pick
list.
4 Click OK. Follow the direction displayed for each step.
A status report window displays the results of the cable calibration.
Calibrating TX Cables Using a Signal Generator and Spectrum Analyzer
Follow the procedure in Table 3-29 to calibrate the TX cables using a
signal generator and spectrum analyzer. Refer to Figure 3-20 for a
diagram of the signal generator and spectrum analyzer.
Table 3-29: Calibrating TX Cables Using Signal Generator and Spectrum Analyzer
Step Action
1 Connect a short test cable between the spectrum analyzer and the signal generator.
2 Set signal generator to 0 dBm at the customer frequency of:
- 869-894 MHz for 800 MHz CDMA
- 1930-1990 MHz for North American PCS.
- 1840-1870 MHz for Korean PCS
3 Use a spectrum analyzer to measure signal generator output (see Figure 3-20, A) and record the
value.
4 Connect the spectrum analyzer’s short cable to point B, (as shown in the lower right portion of the
diagram) to measure cable output at customer frequency of:
- 869-894 MHz for 800 MHz CDMA
- 1930-1990 MHz for North American PCS.
- 1840-1870 MHz for Korean PCS
Record the value at point B.
5 Calibration factor = A - B
Example: Cal = -1 dBm - (-53.5 dBm) = 52.5 dB
NOTE
The short cable is used for calibration only. It is not part of the final test setup. After calibration is
completed, do not re-arrange any cables. Use the equipment setup, as is, to ensure test procedures
use the correct calibration factor.
3