User's Manual

Table Of Contents
Test Set Calibration68P09255A61-4
Aug 2002
1X SC 4812T BTS Optimization/ATP Software Release 2.16.1.x
FIELD TRIAL
3-61
Agilent E4406A Transmitter Tester Self-alignment (Calibration)
The Agilent E4406A requires pre-calibration actions or calibration
verification that are not supported by the LMF. Follow the procedure in
Table 3-27 and refer to Figure 3-19 to perform the Agilent E4406A
self-alignment (calibration).
Table 3-27: Agilent E4406A Self-alignment (Calibration)
Step Action
1 In the SYSTEM section of the instrument front panel, press the System key.
- The softkey labels displayed on the right side of the instrument screen will change.
2 Press the Alignments softkey button to the right of the instrument screen.
- The softkey labels will change.
3 Press the Align All Now softkey button.
- All other instrument functions will be suspended during the alignment.
- The display will change to show progress and results of the alignments performed.
- The alignment will take less than one minute.
Figure 3-19: Agilent E4406A Self-alignment
System
Key
Softkey
Buttons
Softkey Label
Display Area
ti-CDMA-WP-00080-v01-ildoc-ftw
3