User's Manual
GPIB Addresses – continued
JAN 2002
7-49
SC300 1X BTS Hardware Installation, ATP, and FRU Procedures
DRAFT
Verify and Set Agilent E4432B
Signal Generator GPIB
Address
Refer to Figure 7-26 and follow the procedure in Table 7-28 to verify
and, if necessary, change the Agilent E4432B GPIB address.
Table 7-28: Verify and Change Agilent E4432B GPIB Address
Step Action
1 In the MENUS section of the instrument front panel, press the Utility key.
– The softkey labels displayed on the right side of the instrument screen will change.
2 Press the GPIB/RS232 softkey button to the right of the instrument screen.
– The softkey labels will change.
– The current instrument GPIB address will be display below the GPIB Address softkey label.
3 If the current GPIB address is not set to 1, perform the following to change it:
3a Press the GPIB Address softkey button.
– The GPIB Address label and current GPIB address will change to boldface.
– In the on–screen Active Entry Area, Address: and the current GPIB address will display.
3b
On the front panel Numeric keypad, enter the signal generator GPIB address of 1.
– The GPIB Address label will change to Enter.
– Digits entered with the keypad will replace the current GPIB address in the Active Entry display.
NOTE
To correct an entry, press the backspace key at the lower right of the keypad to delete one character at
a time.
3c Press the Enter softkey button to set the new GPIB address.
– The new GPIB address will be display under the GPIB Address softkey label.
7