User's Manual Part 2
Test Set Calibration – continued
DRAFT
Mar 2001
3-57
SCt4812T CDMA BTS Optimization/ATP
Calibrating Test Equipment
The calibrate test equipment function zeros the power measurement level
of the test equipment item that is to be used for TX calibration and audit.
If both a power meter and an analyzer are connected, only the power
meter is zeroed.
Use the Calibrate Test Equipment menu item from the Util menu to
calibrate test equipment. The test equipment must be selected before
calibration can begin. Follow the procedure in Table 3-25 to calibrate the
test equipment.
Prerequisites
Ensure the following prerequisites have been met before proceeding:
Test equipment to be calibrated has been connected correctly for tests
that are to be run.
Test equipment has been selected.
Table 3-25: Test Equipment Calibration
Step Action
1 From the Util menu, select Calibrate Test Equipment.
A Directions window is displayed.
2 Follow the directions provided.
3 Click on Continue to close the Directions window.
A status report window is displayed.
4 Click on OK to close the status report window.
Calibrating Cables
The cable calibration function measures the loss (in dB) for the TX and
RX cables that are to be used for testing. A CDMA analyzer is used to
measure the loss of each cable configuration (TX cable configuration and
RX cable configuration). The cable calibration consists of the following:
Measuring the loss of a short cable – This is required to compensate
for any measurement error of the analyzer. The short cable (used only
for the calibration process) is used in series with both the TX and RX
cable configuration when measuring. The measured loss of the short
cable is deducted from the measured loss of the TX and RX cable
configuration to determine the actual loss of the TX and RX cable
configurations. The result is then adjusted out of both the TX and RX
measurements to compensate for the measured loss.
The short cable plus the RX cable configuration loss is measured –
The RX cable configuration normally consists only of a coax cable
with type-N connectors that is long enough to reach from the BTS RX
port of the test equipment.
. . . continued on next page
3