Installation Instructions

Test Equipment Setup – continued
DEC 2000
7-33
SCt300 BTS Hardware Installation, ATP and FRU Procedures
PRELIMINARY
R3465 Calibration/GPIB
Address & Clock Setup
Follow the steps in Table 7-20 to configure and calibrate the
communication test set.
Table 7-19: R3465 Calibration Procedure
Step Action
NOTE
The calibration should be performed only after the analyzer has been allowed to warm–up and stabilize
for a minimum of 60 minutes.
1 Perform the following to set the standard parameters on the test set:
S Push the SHIFT and PRESET pushbutton (just below the CRT display)
S Select TRANSIENT
Select Setup STD in CRT menu.
2 Verify and if required set the following parameters in the display (rotate the vernier knob to navigate
the screen, push the vernier knob to select):
S Set Type to Japan (800 MHz)
S Set Link to FORWARD
S Verify Rate to 9600/14400 (set Link to REVERSE if Rate needs to be set)
S Set Offset Level to 0.0 dB
S Set Freq Input to FREQUENCY
S Set Cont. Auto Level Set to OFF
Select Return in CRT menu.
3 Verify that the Communications test set GPIB address and Date/Time are set correctly.
S Communications test set GPIB address = 18 (perform the following to view/set as required).
S Select CW
S Push the LCL pushbutton (just below the CRT display)
Push the GPIB and Others CRT menu key to view the current address.
If required, change GPIB address to 18
(rotate the vernier knob to set, push the vernier knob to enter)
S Verify the current Date and Time in upper/right of CRT display
(perform the following to set if required).
Push the Date/Time CRT menu key
If required, change to correct Date/Time
(rotate the vernier knob to select and set, push the vernier knob to enter)
S Select CW