Troubleshooting
None of the channels give proper readings or no readings at all.
Check that the ”Data Transmit” indicator blinks once every 90 seconds (or whatever the transmit
interval is). If it’s not blinking, check that the battery is properly installed and seated, and that
the polarity is correct. Also ensure that the battery is not dead. Check the data acquisition system
for correct settings. Verify that correct transmitter ID number is configured to the data
acquisition system. Check that at least one sensor is connected.
Some of the channels are not giving proper readings.
Data is not sent for channels that do not have sensor connected to them, or if the sensor cable is
detached or cut (in this case also the ”Sensor Fault” blinks). Connect working sensor to the
channel.
Readings seem wrong or inaccurate.
Make sure that correct thermocouple type is selected in the device settings. Check that the device
is not in the direct sunlight.