User Guide
5
Settings
Period
This setting defines the transmit interval. Allowed range is 5...255 seconds. Actual transmit interval
depends on, to a certain degree, the ambient temperature. Default setting is 90 seconds.
This setting has direct impact on the battery life. Avoid using unnecessarily short transmit intervals as
this causes shortened battery life and also limits the number of transmitters that can be used in the
same coverage area. For more information on the effect of transmit interval to battery life, refer to
chapter
Battery.
Inputs
This setting defines the sensor type to be used.
Value
Description
TcB
Linearization for Type
B
thermocouples
TcC Linearization for Type C thermocouples
TcD Linearization for Type D thermocouples
TcE Linearization for Type E thermocouples
TcG
Linearization for Type
G
thermocouples
TcJ Linearization for Type J thermocouples
TcK Linearization for Type K thermocouples
TcL Linearization for Type L thermocouples
TcN
Linearization for Type
N
thermocouples
TcR Linearization for Type R thermocouples
TcS Linearization for Type S thermocouples
TcT Linearization for Type T thermocouples
mV Millivolt measurement, input range -30…2000 mV
Finalizing
Put the lid back in place and tighten the plastic screws. Do not overtighten!
Install the sensors’ ends to their measurement targets.
Configure the ID numbers to the data acquisition system and start the system.
After 90 seconds the measurement results should start appearing.