FTR264-TCK Connecting
Connect K-thermocouple sensor or thermocouple cable equipped with FMTC-K-M type
connector to the cables belonging to the device.
FTR264 Connecting
Thermocouple factory setting is K if you want to change it see the next paragraph.
Open the FTR264 lid by loosening the four plastic screws with a PH2 screwdriver.
Place the lid upside down on the table so you can read the connection sticker on it.
Route the sensor cables through the cable glands.
Detach the screw-post connector and attach the sensor wires according to the sticker’s
instructions, fasten the wires by tightening the screws.
Reattach the screw-post connector securely.
Test the wire connections by gently pulling the wires to see that they are properly fastened.
Tighten the cable glands to ensure the sealing of the enclosure.
Changing the settings
If other than type K thermocouples are used or if the default 90 second transmit interval needs to be
changed, read this chapter and also chapter
Settings, otherwise jump to chapter Finalizing.
Connect a DCS772 USB-POL converter to a USB port of a Windows computer. Attach a POL-
3PIN adapter to the connection cable of the DCS772.
Connect the POL-3PIN adapter to pin-header J3 on the FTR264 circuit board.
Search www.nokeval.com pages with search term MekuWin and download the program.
Install and start MekuWin.
Select settings Protocol SCL, Baud 9600, Parity 8N1, Address 0 and click on Connect Direct.
MekuWin connects to the device. Make the necessary changes to settings and save them.
Detach the cables.