94
COOLSCAN IV ED (LS-40 ED)
Film • 35 mm film: Positive and negative film, both color and mono-
chrome, may be used. Film may be scanned in
strips of up to six frames or mounted as 35 mm
slides using mounts of from 1.0 to 3.2 mm in
thickness and 49.0–50.8 mm in width.
• APS (IX240) film: APS film cartridges can be scanned with the op-
tional IA-20 (S) adapter for APS (IX240) film.
• Prepared slides: Prepared slides can be scanned using the op-
tional FH-G1 slide holder for the MA-20 (S).
Aperture 25.1 × 36.8 mm when scanning 35 mm film in the MA-20 (S)
25.1 × 38.0 mm when scanning 35 mm film in the SA-21
18.6 × 28.4 mm when scanning APS (IX240) film
Scan range 2870 × 4203 pixels when scanning 35 mm film in the MA-20 (S)
2870 × 4332 pixels when scanning 35 mm film in the SA-21
2120 × 3237 pixels when scanning APS (IX240) film
Effective aperture 22.5 × 34.3 mm when scanning 35 mm film in the MA-20 (S)
23.3 × 36.0 mm when scanning 35 mm film in the SA-21
16.1 × 26.9 mm when scanning APS (IX240) film
Effective scan range 2565 × 3911 pixels when scanning 35 mm film in the MA-20 (S)
2657 × 4104 pixels when scanning 35 mm film in the SA-21
1836 × 3067 pixels when scanning APS (IX240) film
Scanning system Fixed-film, movable-plane single-pass optical scanning system
Light source Red, green, and blue (RGB) LEDs
Sensor 2870-pixel, linear CCD image sensor
Color separation Performed by RGB color LEDs
Optical resolution Maximum 2900 pixels per inch
Analog-to-digital (A/D) conversion
12 bits per color (RGB)
Output Full color (8 or 16 bits per color [RGB])
Focus Autofocus
Scanning time • 35 mm film: Approximately 34 s
(Tests performed by Nikon at a resolution of
2900 dpi. Does not include data transfer or dis-
play time.)
• APS (IX240 film): Approximately 26 s
(Tests performed by Nikon at a resolution of
2900 dpi. Does not include data transfer or dis-
play time.)