Product Manual
Table Of Contents
- Nikon XF Series Total Station Instruction Manual
- Legal Notices
- Safety
- Introduction
- Preparation
- Unpacking and Packing the Instrument
- Charging the Battery Pack
- Detaching and Re-Attaching the Battery Pack
- Setting Up the Tripod
- Centering
- Leveling
- Sighting and Focusing
- Setting the Measurement Mode and Preparing the Target
- Measurement in Reflectorless mode
- Setting Up the Prism Reflector
- External Device Connector
- Getting Started
- Checking and Adjustment
- System Diagrams
Nikon XF Series Total Station Instruction Manual 13
Preparation 2
• By design Li-Ion batteries should not be charged above 40 °C-45 °C so a
blinking charge light may mean the batteries are too hot for charging. Charging
will resume after the batteries cool down. The charging time will be longer due
to the batteries cool down when charging batteries above 40 °C-45 °C.
• If the charge indicator(s) are blinking and the batteries feel cool, it may indicate
a problem with the battery or the charger. If the charge light is still blinking
after trying several batteries which are not warm, there is a problem with the
unit or the batteries themselves.
Conditioning / calibrating a battery
• Battery calibration is necessary about once every 6 months or more often if
desired. Calibration ensures the reported battery charge remaining is accurate.
• Hold down the calibration button on the unit and then insert a battery while
holding the calibration button to begin a battery calibration. Only the battery
which was inserted while the button was depressed will begin calibration.
During a battery calibration the battery will be charged, discharged completely,
and then recharged before finishing. Calibration should complete in roughly 22
hours and the charger vents should not be covered during a calibration cycle.
• The blue calibration indicator light(s) will blink slowly (on 1.5 sec, off 2 sec)
while a calibration is in progress and the charge light(s) may be on or off
during the calibration cycle.
• When a calibration cycle is completed, the calibration light will stop blinking
and remain on until the corresponding battery is removed.
• The bottom case temperature may continue to climb up to approximately 43 °C
before temperature regulation is enabled to keep the case from getting warmer.
As the battery voltage drops, the case will cool down and the automatic
temperature limiting will no longer be necessary which minimizes the time it
takes to discharge a battery.
• If the case temperature continues to get too hot internally even after
temperature regulation is enabled, there is a secondary failsafe which will abort
the calibration completely. If an abort occurs, the calibration light(s) will blink
rapidly and battery charging will be re-enabled.










