M306E 04.10.NF.6 (1/2) DS Camera Control Unit DS-L1 DS Camera Head DS-5M DS Cooled Camera Head DS-5Mc < Instructions > (Ver. 2.
Thank you for purchasing the Nikon products. This instruction manual has been prepared for users of the DS Camera Control Unit DS-L1, DS Camera Head DS-5M, and DS Cooled Camera Head DS-5Mc. To ensure correct use, please read this manual carefully before using these products. • The contents of this manual may not be reproduced in whole or in part without permission. • The contents of this manual are subject to change without prior notice.
1. Using the equipment This equipment should be used for photographic purposes only. Do not attempt to use it for any other purpose. 2. Do not disassemble or repair the equipment The DS-L1 and DS AC adapter contain high-voltage power supplies. Never attempt to disassemble or repair the equipment by yourself; doing so may result in electric shock or equipment failure. Never attempt to disassemble any part of the equipment unless instructed to do so in this manual.
1. Installation environment • • • Do not use this equipment in locations subject to high temperatures, high humidity, vibration, or excessive amounts of dust. Doing so may result in fire or malfunction. When used for an extended period of time, the DS-L1 may become hot. Do not place the DS-L1 on a surface that cannot withstand heat (such as vinyl or plastic). Always install the equipment with a surrounding clear area of 100 mm or more.
CONTENTS CONTENTS CONTENTS .................................................................................................... iv I Components of the Equipment ................................................................... 1 II Peripheral Equipment................................................................................ 2 III Names of Parts and Their Functions ............................................................ 5 1 DS Camera Control Unit DS-L1 ...................................
CONTENTS 2.3 Outputting Images Directly to a Printer .......................................................60 2.3.1 2.3.2 Printing an Image ...............................................................................60 Various Print Related Settings ...............................................................60 3 Measuring Two-point Distance or Entering Information—TOOL Menu .............. 62 3.1 Outline of the TOOL Menu .........................................................................
CONTENTS 5.3.1 5.3.2 5.3.3 5.3.4 5.3.5 Installing the Device Driver ................................................................ 113 Outline of the Disk Drive .................................................................... 114 Using the Media Drive ........................................................................ 116 Using the CCD Image Drive ................................................................ 117 Removing the DS-L1 from the PC .................................................
I Components of the Equipment I Components of the Equipment Check to see that all the items listed below are provided in the package. If any items are missing, contact your nearest Nikon representative immediately.
II Peripheral Equipment II Peripheral Equipment (1) Analog RGB input display Although the DS-L1 contains a 1,024 × 768 (XGA) 6.3-inch liquid crystal display, you may need an external display if you want to observe images on a larger screen or at a higher 1,280 × 1,024 resolution (SXGA). Prepare an SXGA 60 Hz or XGA 60 Hz TTL-synchronized display available for use with PCs. For best results, we recommend SXGA monitors.
II Peripheral Equipment (4) USB hub You'll need a USB hub to use a mouse, keyboard, and printer at the same time. Use a USB hub recommended by Nikon. (Not every USB hub is compatible with the DS-L1.) Only one hub can be connected to the DS-L1. If your keyboard is equipped with a USB hub, use this hub. (5) USB printer Use the specified Mitsubishi CP900D or a printer which supports PictBridge Standard. You must purchase a license to use the printer.
II Peripheral Equipment (10) Personal computer (PC) A PC is required for remote control of the DS-L1 or to handle image data on other media. A PC can be connected to the DS-L1 via USB or over a network. For connect a PC to a network, we recommend the following configuration: OS: Memory: Browser: Windows 98/NT/2000/XP Macintosh OS 8.6/9.2/X 64 MB or more Windows Internet Explorer 5.5/6.0 Netscape 6.2/7.0 Macintosh Netscape 6.2/7.
III Names of Parts and Their Functions 1 DS Camera Control Unit DS-L1 III Names of Parts and Their Functions 1 DS Camera Control Unit DS-L1 Front of the DS-L1 Internal LCD 1,024 × 768 dots 6.3-inch LCD display MENU button Press the MENU button to bring up a menu. After a menu is displayed, the MENU button can be used to confirm a selected menu command, just like the ENTER button.
III Names of Parts and Their Functions 1 DS Camera Control Unit DS-L1 Left side of the DS-L1 USB device port Connect a PC to this port with a USB cable to send and receive data to and from the PC. Or connect a PictBridge-supporting printer to perform the direct print. 12 V DC output USB host port This output is Use this port to connect a USB mouse, USB currently unused. keyboard, USB printer (CP900D), or Nikon 12 V DC input 80i/90i microscope. Use a dedicated AC Network Tx indicator adapter.
III Names of Parts and Their Functions 1 DS Camera Control Unit DS-L1 Rear of the DS-L1 RGB connector Camera connector Use this connector to Connect the DS Camera Head DS-5M or the DS Cooled connect an external Camera Head DS-5Mc to this connector. monitor. * Always be sure to turn off the power before connecting or disconnecting the camera head to or from this connector. Failure to do so may result in malfunction.
III Names of Parts and Their Functions 2 DS Camera Head DS-5M / DS Cooled Camera Head DS-5Mc DS Camera Head DS-5M / DS Cooled Camera Head DS-5Mc 2 Tripod mount This mount can be used to mount the camera head on a tripod, etc. Rear Top Bottom C mount cap CAMERA OUT connector Protects the C mount from dust. C mount When taking microphotographs, mount the C mount adapter here before connecting to the microscope. When taking other photographs, attach the C mount lens here.
IV Connecting and Installing the Equipment 1 Connection Diagrams IV Connecting and Installing the Equipment 1 Connection Diagrams Schematic diagram DS-L1 PC or a PictBridgesupporting printer USB cable USB equipment, mouse, keyboard, printer, and Nikon 80i/90i misroscope LAN cable CF card DS AC adapter RGB cable DS camera cable LAN DS-RC DS-5M/DS-5Mc -9-
IV Connecting and Installing the Equipment 1 Connection Diagrams Connectors · DS-L1 Left side PC or a PictBridgeUSB mouse supporting USB keyboard printer ← USB cable USB equipment Printer Microscope DS AC adapter LAN cable LAN ← · DS-L1 Rear RGB cable Remote cable ↓ ↓ ↓ External monitor DS-RC DS-5M/DS-5Mc - 10 - Remote cable
IV Connecting and Installing the Equipment 2 Installing the DS-L1 2 Installing the DS-L1 When installing the DS-L1 on a flat surface such as a desktop by using the stand arm Extract the stand arm from the DS-L1 and place it at an angle of 5 to 10 degrees, as shown below. Leave a clearance of about 100 mm around the DS-L1 to prevent heat from collecting near the equipment.
IV Connecting and Installing the Equipment 3 Connection Methods Connection Methods 3 (1) Connecting the DS Camera Head DS-5M (Equivalent to that for DS-5Mc) Be sure to turn off the power for the DS-L1 before making any connections. Connect one end of the camera cable supplied with the unit to the CAMERA connector on the DS-L1 and the other end to the CAMERA OUT connector on the DS-5M. Squeeze the locking tabs on the connector as you plug in the connector.
IV Connecting and Installing the Equipment 3 Connection Methods (3) Connecting USB Peripheral Equipment Be sure to turn off the power for the DS-L1 before making any connections. When using a USB mouse, USB keyboard, USB printer (Mitsubishi CP900D), or Nikon 80i/90i microscope independently, connect the device directly to the USB-H connector on the DS-L1. When using USB devices simultaneously, connect them to a USB hub. Only one USB hub can be connected to the DS-L1.
IV Connecting and Installing the Equipment 3 Connection Methods Connecting to a PC without going through a LAN Left side of the DS-L1 Third-party 10/100BASE-TX (category 5) cable (crossed) To Ethernet connector PC (7) Connecting External I/O Devices Be sure to turn off the power to the DS-L1 and any equipment to be connected before making connections.
IV Connecting and Installing the Equipment 3 Connection Methods EXTERNAL OUT Functions: Extended external output pins 1 and 2, both for strobe synchronizing output This output is used mainly for synchronized firing of a stroboscope when capturing an image with the DS-L1. Electrical characteristics of strobe synchronizing output are open-collector output, maximum input current = 10 mA, withstand voltage = 24 V, and internal resistance = 100 Ω. The output pulse width is 67 ms.
V Microscope Adjustment V Microscope Adjustment When the camera head is mounted on a microscope, adjust the microscope as described below. (1) Use the Appropriate Filters When capturing a color image: Insert the NCB filter into the optical path. When capturing a black & white image: Insert the appropriate filter into the optical path, depending on the photographic purpose. Generally, inserting a green interference filter (GIF) will heighten contrast.
V Microscope Adjustment (2) Adjusting the Field Diaphragm Adjust the field diaphragm so that it circumscribes the viewfield. The field diaphragm has a significant impact on contrast, especially for fluorescent specimens against dark backgrounds. Make sure you adjust the field diaphragm correctly. For details on adjustment, refer to your microscope’s instruction manual.
V Microscope Adjustment (6) Adjusting the Exposure Time Use an ND filter to adjust the illumination for the microscope to a level that yields a suitable exposure time (generally 1/15 to 1/250 s). You can also adjust the camera sensitivity. Terms What is an ND filter? An ND filter is a filter that affects only the amount of light passed, not the color balance of the light.
SETUP VI Basic Operations 1 Preparations Before Photographing VI Basic Operations 1 Preparations Before Photographing 1.1 Using Various Input Devices 1.1.1 Using the DS-L1 Panel Switches MENU button Cursor buttons ZOOM button FREEZE button CAPTURE button MENU button (menu/enter button) Pressing the MENU button brings up the CAM Easy menu. While the menu is displayed, use this button as the ENTER button to confirm your selection.
.SETUP. VI Basic Operations 1 Preparations Before Photographing Cursor buttons Use these buttons to select an icon in a menu or to increment or decrement a set value. • For an icon to increment or decrement a value such as the Gain icon on the CAM menu Position the cursor over the icon and press the MENU button to select the icon. The color of the cursor frame changes to blue. Press the right or top cursor button to increment the value or the left or bottom cursor button to decrement the value.
VI Basic Operations 1 1.1.3 SETUP Preparations Before Photographing Using a Mouse You can use a USB mouse with the DS-L1. Be sure to use a recommended mouse. Not every USB mouse is compatible with the DS-L1. While most normal operations involve left-clicking the mouse, some may require right-clicking. For icons to set a value or select a mode, left- or right-clicking the mouse button allows you to increment or decrement a value or select items in the forward or reverse direction.
.SETUP. VI Basic Operations 1 1.1.4 Preparations Before Photographing Using a Keyboard You can use a USB keyboard with the DS-L1. For frequent text or character entry, you'll find using a keyboard most convenient. Be sure to use a recommended type of keyboard. Not all USB keyboards are compatible with the DS-L1. Alphanumeric characters in upper- or lowercase can be used with the DS-L1.
VI Basic Operations 1 1.2 SETUP Preparations Before Photographing Notes on Using a CF Card Images can be saved to or reproduced from a CF card. Pay attention to the following when using a CF card: • Use a Type-I or Type II-compliant CF card or CF+ Type II microdrive (1 GB). • Carefully read the user's manual provided with your CF card before using the card. • To format a CF card on your computer, use FAT12, FAT16 (Sometimes these are expressed simply as FAT.), or FAT32.
.SETUP. VI Basic Operations 1 Preparations Before Photographing 1.3 Menu Configuration and Method of Displaying a Menu 1.3.1 Menu Configuration The menus available with the DS-L1 are listed below. Shortcut menus CAM (Easy / Advn. / Reg.
VI Basic Operations 1 1.3.2 SETUP Preparations Before Photographing Method of Displaying a Menu Follow the procedure described below to show or hide a menu or to switch between menus. Using the MENU button on the main unit front panel Press the MENU button to display the CAM Easy menu. To change menus, select your desired tag with one of the cursor buttons and press the MENU button (which functions as the enter button) to confirm. To close the menu screen, press the MENU button once again.
.SETUP. VI Basic Operations 1 1.4 Preparations Before Photographing Initial Settings When using the DS-L1 for the first time after purchasing, you must make the following environment settings before photographing any image. (1) Date and power-save settings Set the unit clock and power-save timer. (2) Image file-save related settings Set an image file format and compression rate, the folder and file names in which to save the file, and the source (photographing mode).
VI Basic Operations 1 SETUP Preparations Before Photographing Date:, Time: Enter a date and time. NOTE: If the DS-L1 remains turned off for extended periods, its internal timer may slow down or stop. When the DS-L1 starts up, it displays the date and time set in its timer (flashes for about 8 seconds). If you have shut off the DS-L1 for one week or longer, check the date and time shown on the DS-L1 when you turn it back on.
.SETUP. VI Basic Operations 1 1.4.2 Preparations Before Photographing Image File Settings—File Settings Display a menu and select the [SETUP] tag on it to enter the SETUP menu. Select File Settings from the SETUP menu to display the File Settings screen shown below. Sets a folder name in which files are to be stored. Sets a file name. Enables or disables the log input function. Sets a file format. When you have finished setting each item, press SAVE press Default to restore data to default values.
SETUP VI Basic Operations 1 Preparations Before Photographing Log If you select the check box labeled “with LOG.txt,” the function for attaching a log (comment statement) is enabled when you save an image. Writing information such as subject name, number, and objective magnification in a log may facilitate future confirmation. You can write up to 100 characters to a log when saving an image to a card or server, or up to 32 characters when outputting to a printer.
.SETUP. VI Basic Operations 1 640*480 Preparations Before Photographing 900KB 450KB 220KB 150KB 120KB When you use an overlay feature on the TOOL menu, it is possible that not all character information may be read, depending on the file type. If information is written in small characters and the image is saved in other than 1280*960 bmp format, some characters may be dropped. Use the appropriate file format when saving an image, according to the table shown below.
VI Basic Operations 1 SETUP Preparations Before Photographing NOTE: To avoid shutter delays when photographing dark subjects, select a combination marked with the performance “Rsp” in the above table. Shutter delay is increased if you save an image in display mode “C.Scan” or select “s5M” for the source in display mode “1.3Mi” or “1.3Mp.” Auto change CCD mode If you set the source to “s5M” and select “Auto change CCD mode,” the display mode is automatically changed between 1.3Mi or 1.
.SETUP. VI Basic Operations 1 1.4.3 Preparations Before Photographing Startup Condition Settings—Power On Settings Display a menu and select the [SETUP] tag on it to enter the SETUP menu. Select Power On Settings from the SETUP menu to display the Power On Settings screen shown below. The following items can be set in the Power On Settings screen. Sets custom mode to be used at startup. When you have finished setting each item, press SAVE press Default to restore data to default values.
VI Basic Operations 1 SETUP Preparations Before Photographing SXGA/XGA,LCD Select a monitor output state from those listed below. XGA (LCD ON) LCD Color: Internal monitor is ON, and external monitor XGA output is ON. (Color adjustment for the internal LCD monitor) SXGA (LCD OFF): Internal monitor is OFF, and external monitor SXGA output is ON. (Color adjustment for the external LCD monitor) XGA (LCD OFF): Internal monitor is OFF, and external monitor XGA output is ON.
.SETUP. VI Basic Operations 1 Preparations Before Photographing "XGA (LCD ON) LCD Color" is initially selected for Power On Settings. When using an external monitor, select "SXGA (LCD OFF)," "XGA (LCD OFF)," or "XGA (LCD ON)." 4) Colors may not display correctly, depending on the external monitor connected to the system.
VI Basic Operations SETUP Preparations Before Photographing NOTE: If the message “NG!! Code Error!!” is displayed when you enter the license code, check to see that the license code you've entered is correct before retrying. The printers to perform the direct printing can be selected from either Mitsubishi CP900D or anyone which supports PictBridge Standard (*).
.CAMERA. VI Basic Operations 2 Photographing Images Photographing Images 2 Adjusting Camera Operation and Exposure for Better Picture Quality—CAM Menu 2.1 The CAM menu is displayed immediately after you display any menu or when you select the CAM tag. The CAM menu has the following functions for setting photographing conditions. 2.1.1 • Four display modes are provided that are classified by resolution, sensitivity, live speed, and balance.
CAMERA VI Basic Operations 2 Photographing Images Photographing procedure To photograph an image from the Easy menu, follow the procedure described below. (1) Choose scene mode and custom mode. Choose these modes according to the subject and the method of observation. (2) Adjust the white balance. Prepare an evenly white subject and adjust the white balance with WB . (3) Adjust the focus. Adjust the image composition and the focus. (4) Correct exposure. Adjust the brightness of the image with CMP.
.CAMERA. VI Basic Operations 2 Photographing Images (1) Choosing scene mode and custom mode Choose a scene mode that suits the subject and the method of observation. Scene mode To change the scene mode for industrial or biological specimens, use Cam OSD on Power On Settings as selected from the SETUP menu. • Industrial specimen scene mode: Ind. • Biological specimen scene mode: Bio. NOTE: The scene mode setup values can also be set and saved as desired.
VI Basic Operations 2 CAMERA Photographing Images Custom modes Custom mode is the mode in which the set values for individual subjects and objectives and those of scene modes are saved. Pressing one of the buttons c1 through c7 calls up the saved content of the selected custom mode. You also can confirm the information in the Info box. NOTE: The method for setting the contents of custom modes are detailed in the explanation of the Advn. menu.
.CAMERA. VI Basic Operations 2 Photographing Images (6) Saving the image Use (Capt. button) or the CAPTURE button on the DS-L1 to save the image. (Capt. button) and the CAPTURE button on the DS-L1 are functionally the same. The image is saved to the destination that has been assigned to the button. You can also use these buttons to print the image. Look at the LED located by the side of the CAPTURE button on the main unit to confirm the destination to which the image will be saved.
CAMERA VI Basic Operations 2 Photographing Images Confirm the image and, if satisfactory, save the image. Use the (capt. button) or the REC menu or the shortcut menu to save the image. If the brightness needs adjustment, change the exposure condition by using the C_gn (camera gain) adjusting button and the shutter speed adjusting button, then photograph again with the (exposure start button). To return from this status to the live image, press (focus button).
.CAMERA. VI Basic Operations 2 2.1.3 Photographing Images Fine Adjustment before Photographing—Advn. Menu Select the [Advn.] tab on the CAM menu, and the Advn. menu will be displayed. In the Advn. menu you can make various settings of the camera individually to obtain an optimum combination of settings for the intended subject. Configuration of the Advn. (Advanced) menu Displayed when the DS-5Mc is connected.
VI Basic Operations 2 CAMERA Photographing Images Photographing procedure Follow the procedure described below for general photography. By referring to the procedure below, find a method that is most suitable for the purpose of observation or the intended subject as you photograph an image. (1) Choose display mode. (2) Choose exposure mode. (3) Choose photometry mode. (4) Choose tone. (5) Adjust the white balance. (6) Adjust the picture quality. (7) Confirm exposure information. (8) Save the image.
.CAMERA. VI Basic Operations 2 Photographing Images (1) Choosing display mode Display mode here refers to one that is used to determine how an image is displayed on the screen. Choose your desired display mode from “1.3Mp,” “1.3Mi,” “5Mi” or “C.Scan.” Normally choose “1.3Mp.” Feeling or features of camera operation vary with a combination of display mode and the source you use during photographing. Refer to “Outline Guide to Source (Photographing Mode) Settings” in “1.4.
CAMERA VI Basic Operations 2 Photographing Images C.Scan: Observing a subject with smoother movement or quickly adjusting the focus With the upper and lower parts of the screen truncated, an image is displayed at a fast rate of 15 frames per second. • Images cannot be saved while being displayed in this observation mode. • Use of this mode is recommended for relatively bright subjects. • Choose this mode when quick focus response is required.
.CAMERA. VI Basic Operations 2 Photographing Images S_AE: Choose “S_AE” when you fix the electronic shutter while photographing the subject. With the electronic shutter fixed to its set value, the DS-L1 automatically sets the camera sensitivity. • Effective in cases where the brightness of the subject changes frequently while being observed under predetermined conditions and you are concerned with the response of automatic exposure.
VI Basic Operations 2 CAMERA Photographing Images (3) Choosing photometry mode There are two photometry modes: Average photometry Peak-hold photometry Normally, average photometry should be selected. The icon indicates the currently selected photometry mode. Sample settings that suit different uses are described below. Try these settings before determining your own settings.
.CAMERA. VI Basic Operations 2 Photographing Images Features of tones There are seven kinds of tones. Choose an easy-to-observe tone according to the subject. TONE1: Wide dynamic range Use this tone when gradation from darks to brights is required (as for subjects that have highlight reflections). Recommended example: IC chips, part-mounted circuit board, gears, etc. TONE2: Weak contrast Use this tone when you want to weaken the contrast.
CAMERA VI Basic Operations 2 Photographing Images The white balance can also be adjusted from a frozen image. 1) While photographing an evenly white subject, press (freeze button) to freeze the display. 2) Press 3) Press WB while in this state, and the white balance will be automatically adjusted. (freeze button) again to release the frozen display, thus completing your settings. In case of over- or underexposure, the white balance cannot be adjusted precisely.
.CAMERA. VI Basic Operations 2 Photographing Images • B&W This is a monochrome image. Use this effect for monochromatic subjects such as DIC or PH and metal tissue. • Nega With the brightness reversed, this is used as a special effect. • BlueBack Blacks in Nega are turned blue. This is used as a special effect. • Sepia With colors developed in sepia, this is used as a special effect. Shading correction SD OFF / SD10% to SD 50% / SD US1 / SD US2 Correct unevenness in shades on the screen.
VI Basic Operations 2 CAMERA Photographing Images Confirm the effective level of the photographed image from image distribution (with a histogram). The DS-L1 calculates image distribution from a frozen image and displays it as a histogram (HISTOGRAM). This function cannot be used unless you use a mouse. Right-click (freeze button) in the Advn. menu with the mouse, and the DS-L1 will start calculating image distribution, with a histogram displayed at the bottom of the menu.
.CAMERA. VI Basic Operations 2 Photographing Images The image is saved according to the File Settings with regard to the directory name, file name, file format, and compression rate. Refer to File Settings (“1.4.2 Image File Settings-File Settings” in Chapter VI) for details on the format in which images are saved. You also can save or print images from a shortcut menu or the REC menu. Refer to “2.2.1 Saving an Image to a CF Card-REC Menu” or “4.6.
CAMERA VI Basic Operations 2 2.1.4 Photographing Images Special Photography on Advn. Menu —Fluorescent Photographing or Dark Subject Photographing The following settings are recommended for fluorescent photography or dark subject photography. Settings on the CAM Advn. menu Displayed only when the DS-5Mc is connected. 1) Set the white balance under normal lighting conditions before photographing. Adjust R and B Gain as necessary.
.CAMERA. VI Basic Operations 2 Photographing Images Other settings required • In File Settings of the SETUP menu, set the source to “s5M” (high picture quality). • In Additional Settings of the SETUP menu, set External Out to “OFF” (without shutter delay). Check the screen for any abnormality. If no problems are found, save the image. NOTE: If you are photographing a dark subject that requires a long exposure time, use the camera head in a cool place to allow for its temperature to decrease.
CAMERA VI Basic Operations 2 Photographing Images Using a CF Card 2.2 The DS-L1 allows you to save a photographed image to a CF card. You also can reproduce a saved image on the screen from the CF card. 2.2.1 Saving an Image to a CF Card—REC Menu Select the [REC] tag at the top of the CAM menu screen. The REC menu will be displayed. In the REC menu, you can specify a directory name and a file name before saving an image.
.CAMERA. VI Basic Operations 2 Photographing Images In the REC menu, you can specify the photographing source, file format, image size, and compression rate when saving an image. • For the source, choose either “s5M” or “s1.3M.” Choose “s5M” to obtain high picture quality. NOTE: Refer to “Outline Guide to Source (Photographing Mode) Settings” in “1.4.2 Image File Settings-File Settings,” Chapter VI for details about the source. • For the file format, choose either “JPEG” or “BMP.
VI Basic Operations 2 CAMERA Photographing Images Saving an image To save an image to a CF card from the REC menu, follow the procedure described below. 1) Choose “1shot” if you want to save only one image sheet. Choose “Continue” to save multiple image sheets at given intervals. 2) Change the destination or file name to or in which to save, as necessary. If you want to save a log, enter a comment in LOG. 3) Press CF , the image will be saved to a CF card.
.CAMERA. VI Basic Operations 2 2.2.2 Photographing Images Reproducing Images from a CF Card—VIEW Menu Display a menu and select the [VIEW] tag on it to enter the VIEW menu. In the VIEW menu, you can reproduce images on the screen from a CF card. You also can delete image files from the CF card or display a thumbnail. Locate the file name or folder name you want to reproduce at the Z position. Reproducing an image Images can only be reproduced in BMP and JPEG file formats.
VI Basic Operations 2 CAMERA Photographing Images Using the formatting function If you insert a CF card formatted in NTFS, i.e., formats not supported by the DS-L1, the following message will be displayed on the VIEW menu. “Error !! Unacceptable CF Card Format (FAT16) ? Format Press the Format ” button to format the CF card in FAT16 format. NOTE: CF cards formatted in other than NTFS may not be correctly formatted in FAT16 depending on the formats used.
.CAMERA. VI Basic Operations 2 Photographing Images 2.3 Outputting Images Directly to a Printer 2.3.1 Printing an Image When operating from a shortcut menu or the REC menu, use the print icon to print. When selecting CP900D printer or a PictBridge-supporting printer in the Additional Settings screen of the SETUP menu window, a print icon either or respectively, is displayed.
VI Basic Operations 2 CAMERA Photographing Images Selecting a source Use File Settings of the SETUP menu to select a source. Choose the “s5M” source to obtain satisfactory picture quality. The features of each source will be the same as when saving an image. Refer to “1.4.2 Image File Settings-File Settings” in Chapter VI for details on how to set the file format, image size, compression rate, and the source.
.TOOL. VI Basic Operations 3 3 3.1 Measuring Two-point Distance or Entering Information—TOOL Menu Measuring Two-point Distance or Entering Information—TOOL Menu Outline of the TOOL Menu Display a menu and select the [TOOL] tag on it, and the TOOL menu will be displayed. In the TOOL menu, the following functions are available to utilize the photographed image. • In overlay, you can measure the distance between two points, enter a comment, or compare images laid on top of another.
TOOL VI Basic Operations 3 Measuring Two-point Distance or Entering Information—TOOL Menu Configuration of the TOOL menu Clear All Erases overlay information Overlay ON/OFF Displays or hides overlay Pen Pen function Marker Count marker function Text Text input function Two-point distance measurement function Calib. Enters a calibration value s1 ~ s7 Calls calibration settings Measure Measures the distance between two points Scale Displays a scale C.
.TOOL. VI Basic Operations 3 3.2 Measuring Two-point Distance or Entering Information—TOOL Menu What is an Overlay? An overlay is something like a transparent sheet that is placed on top of a photographed image. Text, count markers, pen drawings, measurement lines, or superimposed images can be written on the overlay. The contents written to the overlay can be saved or printed along with the image. Basic method of overlay operation All pen drawings, text input, etc.
TOOL VI Basic Operations 3 Measuring Two-point Distance or Entering Information—TOOL Menu Saving or printing the overlay If you save or print an image while an overlay on it is being displayed, the overlay image is written over the photographed image when it is saved or printed. In this case, the photographed image and the overlay are integrated into a single image when saved, so you cannot erase the overlay from the saved image.
.TOOL. VI Basic Operations 3 Measuring Two-point Distance or Entering Information—TOOL Menu Entering a Comment on the Image 3.3 You can enter a comment on the overlay of an image by using the pen drawing, count marking, or text function. An overlay can be placed on top of a live image. If you do not want the image and a comment to be mismatched, use the freeze button to freeze the image before entering a comment. 3.3.
TOOL VI Basic Operations 3 3.3.2 Measuring Two-point Distance or Entering Information—TOOL Menu Using the Count Marker (to Count the Number of Points) The count marker is a convenient function because it displays numbers in the order in which markers have been attached. Attach a marker to each point on the image, and you can count the number of points easily. NOTE: You need a mouse to use this function. 1) Press Marker . The tool bar shown below will be displayed.
.TOOL. VI Basic Operations 3 3.3.3 Measuring Two-point Distance or Entering Information—TOOL Menu Entering Textual Comments The text function allows you to enter a textual comment in the image. NOTE: Although text can be entered from a keyboard, you need a mouse to perform other operations. 1) Press Text . The tool bar shown below will be displayed. 2) Press the red, blue, green, black, white, or yellow color icon to select the text color.
TOOL VI Basic Operations 3 Measuring Two-point Distance or Entering Information—TOOL Menu Measuring the Distance between Two Points on the Screen 3.4 Once you calibrate the reference length, you will have a simple means of measuring distances. You can save seven calibration values for each objective or measurement condition and reuse saved values by calling them up the next time you perform a measurement. 3.4.
.TOOL. VI Basic Operations 3 8) Press Measuring Two-point Distance or Entering Information—TOOL Menu to return to the previous menu. Any calibration values will be erased unless they have been saved. NOTE: Do not save the calibration value when “E” is checked in the calibration screen. Otherwise, an incorrect calibration value will be saved. < When using the magnification of the objective and relay lenses > 1) Select a number in which you want to save the calibration value from s1 through s7 .
VI Basic Operations 3 3.4.2 TOOL Measuring Two-point Distance or Entering Information—TOOL Menu Calling the Calibration Value To use one of the registered calibration values to perform measurement, you must call up the calibration value following the procedure described below before you can use it. NOTE: Be aware that if you call up a calibration value for any image all of the overlay information for that image will be erased. 1) Select the calibration value you want to use, from s1 through s7 .
.TOOL. VI Basic Operations 3 3) Measuring Two-point Distance or Entering Information—TOOL Menu Draw a measurement line. Click the mouse button at the start point of the object to be measured and hold down the mouse button while you drag and release the mouse button at the end point, to draw a line. The length of the measurement line will be displayed in the vicinity of it. Each measurement line is assigned numbers 1 to 9. The number of measurement lines drawn in the same color is limited to nine.
VI Basic Operations 3 TOOL Measuring Two-point Distance or Entering Information—TOOL Menu 3.5 Adjusting the Subject Position and Comparing Sizes 3.5.1 Using Crosshairs to Adjust the Subject Position You can display crosshairs on the screen and use it to adjust the center or the horizontal or vertical position of the subject. 1) Press Cross , and crosshairs will be displayed. 2) Point to the center of the crosshairs and click the mouse button. Hold down the button while you move the mouse.
.TOOL. VI Basic Operations 3 Measuring Two-point Distance or Entering Information—TOOL Menu 8) “Center Cross” is an option that allows you to fill the center part of the crosshairs. Select the “Yes” check box to fill the center part of the crosshairs. If you select the “No” check box, the center part of the crosshairs will not be filled. 9) Press Center screen. 10) Press Save to save the crosshairs settings.
TOOL VI Basic Operations 3 3.5.2 Measuring Two-point Distance or Entering Information—TOOL Menu Using a Scale to Measure the Subject Size You can display a scale on the screen and use it to measure an approximate size of the subject. 1) Select the calibration value you want to use from s1 through s7 . 2) Press Scale , and a scale matched to the selected calibration value will be displayed. 3) Press Scale again to erase the displayed scale.
.TOOL. VI Basic Operations 3 3.5.3 Measuring Two-point Distance or Entering Information—TOOL Menu Using XY Scales to Measure the Dimensions of the Subject By displaying XY scales on the screen, you can measure the approximate dimensions of the subject. Because the X and Y scales can be moved independently, you can move the scales close to the subject before measuring its dimensions.
VI Basic Operations 3 TOOL Measuring Two-point Distance or Entering Information—TOOL Menu NOTE: * If the microscope connected to your system is the Nikon 80i/90i microscope using a motorized nosepiece or intelligent nosepiece, note that each calibration value call number is linked to the hole number of the nosepiece. This means you cannot change calibration values by selecting s1 through s7 . Calibration values are automatically changed when you change the objective. 3.5.
.TOOL. VI Basic Operations 3 3.5.5 Measuring Two-point Distance or Entering Information—TOOL Menu Using Dual-Window Display to Compare with the Reference Image You can split the screen into left and right halves and display a still image in the left half and a live image in the right half for image comparison. You can use the electronic zoom to zoom in or out in either half of the screen. If you are using a mouse, you can scroll the display up or down on either screen. 1) Select DWIN on the VIEW menu.
VI Basic Operations 4 4 4.1 NETWORK Using the DS-L1 via a Network Using the DS-L1 via a Network Typical Method of Use You can connect the DS-L1 and a PC in a network and operate the DS-L1 from the PC or exchange image files between the DS-L1 and the PC. There are three methods for operating the DS-L1 from a PC: • Use a WEB browser • Use an FTP command • Use a Telnet command You also can access an FTP server PC from the DS-L1 via a network. Possible operations in each method are described below.
.NETWORK. VI Basic Operations 4 Using the DS-L1 via a Network Using FTP commands to operate the DS-L1 from a PC • Get images from the DS-L1 • Send and save images to the DS-L1 • Advanced control of the DS-L1 FTP commands may be used from a PC to receive, send, or save image files. Furthermore, advanced control of the DS-L1 is possible using the DS-L1’s FTP server function.
VI Basic Operations 4 NETWORK Using the DS-L1 via a Network 4.2 Network Settings 4.2.1 Items Needing Confirmation Items Needing Confirmation A LAN in some DS-L1 operating environments may have been set up in a way exclusive to that environment, so be sure to consult the network administrator in your organization before setting up a PC. This is necessary to avoid causing problems with the LAN.
.NETWORK. VI Basic Operations 4 4.2.2 Using the DS-L1 via a Network Setting Essential Items—Network Settings In this instruction manual, explanation are given assuming that the IP address of the DS-L1 is 192.168.10.2, that of the server is 192.168.10.1, and that of the subnet mask is 255.255.0.0., while the gateway address is 192.168.128.1. Of course, you should use the IP address assigned to you by your network administrator.
NETWORK VI Basic Operations 4 Using the DS-L1 via a Network Internal Server Set the IP address, subnet mask, user ID, and password of the DS-L1. Refer to “Settings for improved security” described below for details on how to set a password. • When automatically acquiring an IP address D.H.C.P * ( Auto get an IP ) IP Address Automatically set after restarting Subnet Mask Automatically set after restarting NOTE: The DS-L1 takes some time after it restarts before the addresses are displayed.
.NETWORK. VI Basic Operations 4 Using the DS-L1 via a Network Limiting connectable IPs If you want to limit the IP addresses connectable to the DS-L1, select the “IP Authority” check box. To register the IP addresses that are allowed to connect to the DS-L1, follow the procedure described below. Method for registering the IP addresses that are allowed to connect 1) Insert a CF card into the card slot of your computer. NOTE: A card adapter may be needed depending on your computer.
NETWORK VI Basic Operations 4 Using the DS-L1 via a Network 4.3 Operating the DS-L1 from a WEB Browser 4.3.1 Setting the WEB Browser 4.3.1(1) When Using Internet Explorer 1) Start the PC. Refer to your computer’s user manual for details. 2) Launch Internet Explorer. Refer to your computer’s user manual or the Internet Explorer help function for details. 3) Select [Tools] - [Internet Options]. Left-click here to open the pull-down menu. Left-click here. 4) Select [Connections].
.NETWORK. VI Basic Operations 4 6) Using the DS-L1 via a Network Select [Advanced]. If [Advanced] is disabled and you cannot select it, left-click on [Use a Proxy Server] first. A check mark will be displayed, allowing you to select [Advanced]. Left-click here. 7) In the [Exceptions] column, write the IP address of the DS-L1. (Refer to “4.2.2 Setting Essential Items” in Chapter VI for details on the IP address of the DS-L1.) • Enter a period after every three digits of the IP address.
VI Basic Operations 4 4.3.1(2) NETWORK Using the DS-L1 via a Network When Using Netscape 1) Start the PC. Refer to your computer’s user manual for details. 2) Launch Netscape Refer to your computer’s user manual or the Netscape help function for details. 3) Select [Edit] - [Preferences]. Left-click here to open the pull-down menu and select [Preferences] from the menu. 4) Select [Advanced] - [Proxies]. Lift-click [Advanced]. When a menu appears under [Advanced], left-click on [Proxies].
.NETWORK. VI Basic Operations 4 5) Using the DS-L1 via a Network Select the check box labeled [Manual Proxy Configuration] and then left-click on [View]. Left-click here. Left-click here. 6) In the [Exceptions] column, write the ID address of the DS-L1. (Refer to “4.2.2 Setting Essential Items” in Chapter VI for details on the IP address of the DS-L1.) • Enter a period after every three digits of the IP address. • If the first digits of the IP address are “00” or “0,” do not enter the zeros.
VI Basic Operations 4 4.3.2 NETWORK Using the DS-L1 via a Network Using WEB Screens Once a PC and the DS-L1 are connected through a network, it is possible to check or capture the image that is being shot by the camera head of the DS-L1 from a WEB browser. Even from a distance, you can check what image the DS-L1 is currently shooting. When you access the DS-L1 from a WEB browser, a WEB screen is displayed on which you can operate the DS-L1.
.NETWORK. VI Basic Operations 4 Using the DS-L1 via a Network Opening a WEB screen 1) Launch either Internet Explorer or Netscape Communicator on your computer. 2) Enter the IP address of the DS-L1 in the URL address field. (Example: http://192.168.10.2/) The Entrance screen of the DS-L1 will appear. 3) Click either ROOM1 or ROOM2. Click either ROOM1 or ROOM2. ROOM1 and ROOM2 are explained in more detail in the subsequent pages.
NETWORK VI Basic Operations 4 Using the DS-L1 via a Network Using ROOM1 ROOM1 allows you to observe the image that is being shot by the DS-L1 at specified intervals, while at the same time capturing images onto the PC or displaying a stream window. You also can control the DS-L1 main unit from this screen. (2) (3) (4) (5) (9) (1) (6) (7) (8) (10) (1) Tabs Entrance: Reload: ROOM2: Returns you to the Entrance screen. Redisplays the ROOM1 screen, with the preview image updated.
.NETWORK. VI Basic Operations 4 Using the DS-L1 via a Network (3) Control Pad Displays a subwindow for controlling the DS-L1. From this window, you can save the latest image to an image file. You can also perform almost all operations that you would perform from the CAM Advn. menu. (Refer to “2.1.3 Fine Adjustment before Photographing—Advn. Menu” in Chapter VI.) NOTE: No data or values can be entered directly from the keyboard. Use a mouse to control the buttons on the WEB screen.
VI Basic Operations 4 NETWORK Using the DS-L1 via a Network NOTE: * If you click the [Cancel] button in the folder select screen, it is possible that the browser will no longer be updated. In such a case, wait for a while or temporarily close the browser and reopen it. * Sometimes, the window may be closed while downloading an image file. To avoid this problem, we recommend installing SP1 (Service Pack 1) if you are using Internet Explorer 5.5.
.NETWORK. VI Basic Operations 4 Using the DS-L1 via a Network (6) E.Zoom (7) Direction buttons (8) Sample image E.Zoom allows you to turn the electronic zoom on or off. When you select the E.Zoom check box, the direction buttons in (7) and the sample image in (8) are displayed, with the preview screen displayed in a 4-fold enlarged view. The gray region in the sample image (white rectangle) shows an enlarged view of what is currently being displayed.
NETWORK VI Basic Operations 4 Using the DS-L1 via a Network Using ROOM2 ROOM2 displays a list of image files saved to a CF card. The image files can be downloaded from here to the PC. (4) (3) (2) (1) (5) (1) Tabs Entrance: ROOM1: Reload: Returns you to the Entrance screen. Displays the ROOM1 screen. Redisplays the ROOM2 screen. (2) Location The DS-L1 installation information, etc., that has been set by Telnet commands is displayed here.
.NETWORK. VI Basic Operations 4 Using the DS-L1 via a Network Using FTP Commands to Operate the DS-L1 4.4 You can run FTP commands on your PC to get images from the DS-L1. NOTE: Always be sure to consult the network administrator in your organization before executing FTP commands. Your organization’s network may have been set up in a unique manner, so invalid operations might cause a problem on your network. 4.4.
VI Basic Operations 4 4.5 NETWORK Using the DS-L1 via a Network Using Telnet Commands to Operate on the DS-L1 You can use Telnet commands to operate the DS-L1 from your PC. NOTE: Always be sure to consult the network administrator in your organization before executing Telnet commands. Your organization’s network may have been set up in a unique manner, so invalid operations could cause a problem on your network.
.NETWORK. VI Basic Operations 4 Using the DS-L1 via a Network Saving Images to an FTP Server PC from the DS-L1 4.6 In this instruction manual, explanations are given assuming that the IP address of the DS-L1 is 192.168.10.2, that of the server is 192.168.10.1, and that of the subnet mask is 255.255.0.0., while the gateway address is 192.168.128.1. In practice, you should use the IP address assigned to you by the network administrator in your organization.
VI Basic Operations 4 NETWORK Using the DS-L1 via a Network Settings Adding a user Create a new user account. 1) From the [Start] menu, select [Programs] - [Administrative Tools] - [User Manager]. 2) Enter a user ID and a password. To ensure that the same information has been set for External Server in the Network Settings of the DS-L1, enter the user ID and the password using ten alphanumeric characters in lower- or uppercase.
.NETWORK. VI Basic Operations 4 Using the DS-L1 via a Network When specifying a directory for saving separate image files for multiple DS-L1s that have been connected 1) Select a user in User Manager, and specify a local path for the home directory in [Profile]. (The specified path should be below the home directory, however.) Example: If the home directory is “C:\DSFTP,” the local path in the user account profile of the DS-L1 is “C:\DSFTP\DS1,” and the FTP Dir.
VI Basic Operations 4 NETWORK Using the DS-L1 via a Network Adding components If Internet Information Service (IIS) does not have components added, you need to install them. 1) Open [Control Panel] and select [Add/Remove Programs] - [Add/Remove Windows Components]. 2) Select Internet Information Service (IIS) and click the [Advanced] button. 3) Examine whether the [FTP (File Transfer Protocol) Services] check box is selected. If not, select it and click [OK].
.NETWORK. VI Basic Operations 4 Using the DS-L1 via a Network Other settings When enabling logon as an “anonymous” user In order to discriminate the user by which image files are written to the FTP server, we in the above settings adopted a method to log in with a user ID and a password. However, it is possible to log in as an “anonymous” user. 1) When setting External Server in Network Settings of the DS-L1, select “anonymous” for the User ID and “none” for the Password.
VI Basic Operations 4 4.6.1(3) NETWORK Using the DS-L1 via a Network When Using Mac OS X Setup procedure (1) Setting up the FTP server Add FTP server functions. (2) Adding a user (entering the user ID and password) Set the DS-L1 as a log-in user. This setting must be made after logging-in with the user ID of the system administrator. Settings Setting the FTP server 1) Select [Apple Logo] - [System Environment Settings].
.NETWORK. VI Basic Operations 4 Using the DS-L1 via a Network Other settings Network settings and confirmation Network settings and confirmation can also be accomplished by selecting [Macintosh HD] – [Applications] – [Utilities] – [Network Utility].
NETWORK VI Basic Operations 4 4.6.2 Using the DS-L1 via a Network Setting External Servers of the DS-L1—Network Settings Consult the network administrator in your organization for details on the FTP server. If you are going to connect the DS-L1 to a LAN, you need to enable it for network operation and set its own address. Setting the FTP client function To set the FTP client function of the DS-L1, select Network Settings from the SETUP menu to call up the Network Settings screen shown below.
.NETWORK. VI Basic Operations 4 Using the DS-L1 via a Network Connection test Ping test To test the connection state of the external server, execute Ping Test on the Network Settings screen. This function is used to inspect whether the specified server exists in the LAN. This connection test is performed on the server that is shown in the “IP Address” column of External Server. However, it is necessary as a premise that the IP Address has been saved by pressing SAVE .
VI Basic Operations 4 4.6.3 NETWORK Using the DS-L1 via a Network Saving Image Files in the FTP Server Image files can be saved in the FTP server by using either the CAPTURE button or the REC menu. 4.6.3(1) Using the CAPTURE Button to Save Image Files Saving image files in the FTP server The CAPTURE button on the DS-L1 main unit or the DS Remote Controller DS-RC may be used to save image files in the FTP server.
.NETWORK. VI Basic Operations 4 4.6.3(2) Using the DS-L1 via a Network Using the CAM Menu to Save Image Files Select the [REC] tag at the top of the CAM menu screen to display the REC menu. In the REC menu, you can specify a directory and a file name before saving an image. You can also choose interval photography, which lets you shoot an image repeatedly at given intervals. The following explains how to save images from the REC menu.
VI Basic Operations 4 NETWORK Using the DS-L1 via a Network Saving an image To save an image to the FTP server from the REC menu, follow the procedure described below. 1) Choose “1shot” if you want to save only one sheet image. Choose “Continue” to save multiple sheets of images at given intervals. 2) Change the destination or file name to or in which to save, as necessary. If you want to save a log, enter a comment in LOG. 3) Press FTP , and the image will be saved to the FTP server.
.USB. VI Basic Operations 5 5 5.1 Using the DS-L1 when Connected to a PC by a USB Cable Using the DS-L1 when Connected to a PC by a USB Cable Recommended Operating Environment When the DS-L1 is connected to your PC by a USB cable, you can read or write to the recording media inserted in the DS-L1 or capture the live image being shot by the CCD camera onto the PC. Supported PCs The DS-L1 can only be connected to a Windows-based. However, this is limited to PCs that have a USB interface board installed.
USB VI Basic Operations 5 Using the DS-L1 when Connected to a PC by a USB Cable USB interface boards and hubs The DS-L1 supports USB 2.0. If the DS-L1 and the USB interface board are connected in one of the following ways, data is transferred at a rate of 12 Mbps compliant with the USB 1.1 standard. USB 1.1 interface board DS-L1 USB 1.1 interface board USB 1.1 hub DS-L1 USB 2.0 interface board USB 1.
.USB. VI Basic Operations 5 Using the DS-L1 when Connected to a PC by a USB Cable USB host drivers Use the USB host drivers made by Microsoft Corporation, as the drivers included with your interface board may not operate properly. • When using the USB 1.1 interface board: The driver is included with the OS • When using the USB 2.0 interface board: The driver is included with the OS providing it is Windows XP Service Pack 1 (SP1) or newer.
VI Basic Operations 5 Using the DS-L1 when Connected to a PC by a USB Cable 5.3 Using the DS-L1 in Windows 2000 5.3.1 Installing the Device Driver When the DS-L1 is connected to a PC for the first time, the PC automatically recognizes the DS-L1 and installs the device driver that comes standard with the OS. During the installation process, the window shown below will be displayed sequentially. You do not need to perform any specific operations.
.USB. VI Basic Operations 5 5.3.2 Using the DS-L1 when Connected to a PC by a USB Cable Outline of the Disk Drive When the DS-L1 is connected to a PC, the disk drive for the CF card, etc., and the internal image drive of the DS- L1 are each recognized as one removable disk by the PC. When the DS-L1 is connected to a PC, two removable disks are displayed in [My Computer] on your PC’s desktop. The drive names of the removable disks differ with each PC environment.
VI Basic Operations 5 USB Using the DS-L1 when Connected to a PC by a USB Cable How to discriminate between the media drive and the CCD image drive Right-click any removable disk icon on the My Computer screen to display its [Properties] dialog box. This dialog box helps you identify the drive assigned to that removable disk. In the case of the CCD image drive, for example, it can be identified based on the volume label, “NIKONIMG,” and its capacity, 16.0 MB, as shown below.
.USB. VI Basic Operations 5 5.3.3 Using the DS-L1 when Connected to a PC by a USB Cable Using the Media Drive Sending/receiving image files Insert media into the card slot of the DS-L1 and double-click the media drive (removable disk) icon on the My Computer screen to access the media. The image files on the media drive can be retrieved from the media or transferred from the PC to other media files in the same way as files are normally manipulated on a PC.
USB VI Basic Operations 5 5.3.4 Using the DS-L1 when Connected to a PC by a USB Cable Using the CCD Image Drive Receiving image files Double-click the CCD image drive (removable disk) icon on the My Computer screen to access the internal [DS-L1IMG] folder of the DS-L1. The latest image files shot by the DS-L1 are stored in the [DS-L1IMG] folder, and these files can be downloaded to the PC. NOTE: Because the CCD image drive is read-only, no image files can be sent from the PC to the DS-L1.
.USB. VI Basic Operations 5 Using the DS-L1 when Connected to a PC by a USB Cable Confirming the updated date/time of image files 1) Open My Computer and display image file icons. Then select [View] - [Refresh] to update the image file icons displayed. 2) Right-click any image file icon of interest and select [Properties] from the ensuing menu. The updated date and time will be displayed.
VI Basic Operations 5 5.3.5 USB Using the DS-L1 when Connected to a PC by a USB Cable Removing the DS-L1 from the PC Because the DS-L1 is connected via a USB interface, it can be removed from the PC even if the PC remains powered on. However, to prevent data from being corrupted by removing the DS-L1 during write operation, follow the procedure described below. 1) Eject the media from the media drive before removing the DS-L1. 2) Unplug the USB cable or turn off the power to the DS-L1.
.USB. VI Basic Operations 5 Using the DS-L1 when Connected to a PC by a USB Cable 5.4 Using the DS-L1 in Windows XP 5.4.1 Installing the Device Driver When the DS-L1 is connected to a PC for the first time, the PC automatically recognizes the DS-L1 and installs the device driver that comes standard with the OS.
VI Basic Operations 5 5.4.2 USB Using the DS-L1 when Connected to a PC by a USB Cable Outline of the Disk Drive When the DS-L1 is connected to a PC, the disk drive for the CF card, etc., and the internal image drive of the DS-L1 are each recognized as one removable disk by the PC. A screen may be displayed prompting for confirmation of how you want the removable disk to be handled by the OS. For the media drive, this screen is displayed at the same time media is inserted into the media drive.
.USB. VI Basic Operations 5 5.4.3 Using the DS-L1 when Connected to a PC by a USB Cable Using the Media Drive Sending/receiving image files Insert media into the card slot of the DS-L1 and display the Explorer screen from the confirmation screen described above. You can access media from this screen. Or you can double-click the media drive (removable disk) icon located below “Devices with Removable Storage” on the My Computer screen to access the media.
USB VI Basic Operations 5 5.4.4 Using the DS-L1 when Connected to a PC by a USB Cable Using the CCD Image Drive Double-click the CCD image drive (removable disk) icon on the My Computer screen to access the internal [DS-L1IMG] folder of the DS-L1. The latest image files that have been shot by the DS-L1 are stored in the [DS-L1IMG] folder, and these files can be downloaded to the PC. NOTE: Because the CCD image drive is read-only, no image files can be sent from the PC to the DS-L1.
.USB. VI Basic Operations 5 Using the DS-L1 when Connected to a PC by a USB Cable Confirming the updated date/time of image files 1) Open My Computer and display image file icons. Then select [View] - [Refresh] to update the image file icons displayed. 2) Right-click any image file icon of interest and select [Properties] from the ensuing menu. The updated date and time will be displayed.
VI Basic Operations 5 5.4.5 USB Using the DS-L1 when Connected to a PC by a USB Cable Removing the DS-L1 from the PC Because the DS-L1 is connected via a USB interface, it can be removed from the PC even while the PC remains powered on. However, to prevent data from being corrupted while removing the DS-L1 during a write operation, follow the procedure described below. 1) Eject the media from the media drive before removing the DS-L1. 2) Click the “Safe to Remove Hardware” icon on the task bar.
.MICROSCOPE. VI Basic Operations 6 Using the Microscope Control Using the Microscope Control 6 Overview 6.1 If the DS-L1 and the Nikon 80i/90i microscope are connected via USB, the microscope can be controlled using the DS-L1 menus. The microscope status can also be displayed on DS-L1 menus. The items for which the microscope can be controlled and its status can be displayed this way depends on the specific microscope in use.
MICROSCOPE VI Basic Operations 6 Using the Microscope Control Control the lamp • • Change the condenser • Adjust the diascopic illumination aperture diaphragm • Adjust the diascopic illumination field diaphragm • Adjust the episcopic illumination field diaphragm • Adjust the transmission factor of the ND filter • Change the analyzer (MIC Info menu) • Show the name of the objective • Show the position of the sta
.MICROSCOPE.
MICROSCOPE VI Basic Operations 6 4) Using the Microscope Control When the Nikon 90i microscope incorporating DIH-M is connected Parts Menu displayed Focus (Z axis) Nosepiece Control/status display menu displayed Motorized nosepiece Control/status display menu displayed Optical zoom Status display menu displayed Optical path Status display menu displayed Shutter Control/status display menu displayed Filter cube Status display menu displayed Lamp Control/status display menu displayed Univer
.MICROSCOPE.
MICROSCOPE VI Basic Operations 6 6.2.1 Using the Microscope Control Microscope Control Menu 1 Select the tab [1] of the microscope control menu to display "microscope control menu 1." In "microscope control menu 1," you can control the motorized units of the microscope used often. Here, the menus displayed on DS-L1 are explained for the case in which the Nikon 90i microscope incorporating DIH-E and having all motor-driven parts attached (full spec) is connected in a microscope system.
.MICROSCOPE. VI Basic Operations 6 Using the Microscope Control Focus (Adjusts focus) Compatible only with the Nikon 90i microscope. 1) Set the coarse/fine select icon to coarse motion mode. (The arrow icon changes shape for coarse motion and fine motion modes. for coarse motion mode, for fine motion mode.) 2) Use the Up icon 3) Finally, to fine adjust, set the coarse/fine select icon to fine motion mode and use the Up icon and the Down icon to adjust the focus position.
VI Basic Operations 6 MICROSCOPE Using the Microscope Control NOTE: * If autofocus does not work correctly, try moving the sample until a sharply contrasting part of the sample is at center of the view field. Then reexecute autofocus. * Supported samples are specimen slide with cover glass. (Thickness of the slide glass + cut piece: 0.9 mm to 1.7 mm, both ends inclusive) * Slide glass of thickness of up to 1.2 mm are supported.
.MICROSCOPE. VI Basic Operations 6 Using the Microscope Control NOTE: * When you change the objective, the diascopic illumination aperture diaphragm, diascopic illumination field diaphragm, condenser, Focus and ND filter also are changed accordingly. For more information, refer to the instruction manual for your microscope. * When you change the objective, all the overlay data, excluding the XY scale and the scale currently displayed, is deleted.
VI Basic Operations 6 MICROSCOPE Using the Microscope Control NOTE: * When you are using a microscope incorporating the DIH-M, only the current optical path is displayed. * Each time you left-click to select, the optical path changes sequentially from Binocular to Front port and Rear port and back to Binocular. When you right-click to select, the optical path changes sequentially from Binocular to Rear port and Front port and back to Binocular.
.MICROSCOPE. VI Basic Operations 6 Using the Microscope Control NOTE: * Before this function can be used, you must first set up filter cube information from a PC using the setup software. For more information, consult the instruction manual for your microscope. * If you are using a microscope incorporating the DIH-M, only the filter name is displayed. * Right-click on an icon representing a filter number, and the names of all registered filters will be displayed.
MICROSCOPE VI Basic Operations 6 6.2.2 Using the Microscope Control Microscope Control Menu 2 Select tab [2] of the microscope control menu to display "microscope control menu 2." This section describes the menus displayed on the DS-L1 when the microscope connected is a Nikon 90i microscope incorporating the DIH-E and having all motor-driven parts attached (full spec). If any of these components are missing or manual parts are attached, no menus will be displayed for those components.
.MICROSCOPE. VI Basic Operations 6 Using the Microscope Control Condenser Compatible only with the Nikon 90i microscope. 1) In the condenser column , select a condenser number to which your desired condenser is registered. The condenser will change to the one you select. 2) After switching condensers, confirm that the selected number icon is marked with a red border and that the condenser number and name are displayed.
MICROSCOPE VI Basic Operations 6 Using the Microscope Control NOTE: The level indicator shows the current aperture of the diaphragm in increments of 2.0 in the range 0.8 to 30.6. Episcopic illumination field diaphragm Compatible only with microscopes incorporating the DIH-E. 1) In the episcopic illumination field diaphragm column , select the icon to open the diaphragm in increments of 0.1.
.MICROSCOPE. VI Basic Operations 6 6.2.3 Using the Microscope Control Microscope Status Display/Microscopy Setup Menu (MIC Info Menu) Right-click anywhere on the screen where no other menu is displayed to display the microscope status display/microscopy setup menu (MIC Info menu). NOTE: This menu is displayed only when the connected microscope is a Nikon 90i microscope or Nikon 80i microscope incorporating the DIH-E.
VI Basic Operations 6 MICROSCOPE Using the Microscope Control Objective Compatible only with Nikon 80i/90i microscopes that have motorized or intelligent nosepieces attached. 1) The menu indicates the magnification and NA value of the current objective in the nosepiece column . NOTE: * Before using this function, you must set up the objective information from a PC with the setup software.
.MICROSCOPE. VI Basic Operations 6 Using the Microscope Control Filter Compatible only with microscopes incorporating the DIH. 1) The menu indicates the name of the current filter in the filter column . NOTE: * Before using this function, you must set up filter information from a PC with the setup software. For more information, refer to the instruction manual for your microscope. * Use microscope control menu 1 to change filters.
VI Basic Operations 6 MICROSCOPE Using the Microscope Control 1) Right-click on the icon indicating an observation method to open the menu showing all observation methods. 2) Enter a check mark in the check box.
.MICROSCOPE. VI Basic Operations 6 Using the Microscope Control Autofocus Precautions 6.3 Before using autofocus, check your microscope for the following. Disregarding these precautions may result in damage to the objective or sample. Precautions for samples 1) The autofocus function may be used with samples mounted on specimen slides with cover glass in the sizes indicated below. • Thickness: 0.9 mm to 1.
VI Basic Operations 6 MICROSCOPE Using the Microscope Control Precautions for 90i setup 1) Make sure the observation method is set to bright-field microscopy. 2) Make sure the objective used is • N.A. 0.04 to 0.95 • x1 to x100. (Phase contrast, liquid immersion, polarization, Hoffman, multi-immersion, and industrial-use objectives may not be used.) 3) Make sure the objective information is set at each address on the nosepiece.
.MICROSCOPE. VI Basic Operations 6 Using the Microscope Control ♦ List of autofocus messages The following messages may be displayed while the microscope is autofocusing: Message Now Focusing Content Now autofocusing. Do not alter the microscopy conditions. Complete Autofocusing complete. Comp. Low Contrast Sample lacking sufficient contrast for autofocusing. Slight focal difference may take place in some conditions. Perform manual focusing if this happens. Warn. Exp.
VII Care and Maintenance 1 Cleaning the System VII Care and Maintenance 1 2 Cleaning the System • Use a soft cloth to wipe away dirt from the DS-L1, the remote controller, and the cables. • Before cleaning the camera head, first attach the C mount cap. Then use a soft cloth to wipe away dirt. • Use a blower to blow away dirt inside the camera head mount. When doing so, be sure not to touch the glass filter on the front of the CCD.
VIII Troubleshooting 1 Power Supply VIII Troubleshooting If you experience any problems, check the troubleshooting list below. If you are still not able to resolve the problem, consult with your nearest Nikon representative. Power Supply 1 Symptom The power does not come on. Corrective action • Make sure the power cord is plugged firmly into an AC outlet at one end and the DS AC adapter and then the DC-IN connector on the DS-L1 main unit on the other. • Use the power cord supplied with the system.
VIII Troubleshooting 2 Image Output Image Output 2 Symptom No image appears. Corrective action • Connect the cables securely. • Connect the cables firmly in their specified locations. • Connect the camera head correctly. • Remove the C mount cap from the camera head. • Place the subject within the field of view of the microscope or lens. • Direct the optical path of the microscope to the camera. • Set the exposure correctly. => See “2 Photographing Images” in Chapter VI.
VIII Troubleshooting 2 Image Output Symptom The color is poor. Corrective action • Set the white balance. => See “2 Photographing Images” in Chapter VI. • Apply a moderate amount of color compensation. => See “2 Photographing Images” in Chapter VI. • Turn off hue adjustment. => See “2 Photographing Images” in Chapter VI. • Set the color temperature of the internal monitor and external monitor to an appropriate level. => See “1.4.4 Adjusting Monitor Output for an External Monitor” in Chapter VI.
VIII Troubleshooting 3 Photographing and Saving Photographing and Saving 3 Symptom Captured images have low resolution and the overlay information cannot be read. Photographing is mistimed even when freezing. Camera modes change unintentionally. Corrective action • Set the image quality in File Settings of the SETUP menu. Reduce the JPEG Quality (compression rate) or select “No compression (BMP).” The recommended file settings are BMP in image size of 1280*960. => See “1.4.
VIII Troubleshooting 5 5 Networking Networking Refer to “4.2 Network Settings” in Chapter VI for details about network settings. Symptom Cannot connect the system to a network. => See “4.2 Network Settings” in Chapter VI. Corrective action Follow the sequence shown below to check network settings.
VIII Troubleshooting 5 Networking Symptom Corrective action Cannot connect the Checking for presence in a LAN system to a network. (Continued from the preceding page) Check network connections by using the “ping” command. For Windows, select [Start] - [Programs] - [Accessories] - [MS-DOS Prompt] in that order, and enter the command from the prompt given below.
VIII Troubleshooting 5 Symptom Networking Corrective action Images are When images are sent via FTP, the DS-L1 does not check for duplicate file overwritten when names. saved to an FTP Although file names are also suffixed with a serial number when Auto server. setting is not used, if the power is turned off once, the serial numbers recycle and start from 1 again. Therefore, it is possible that one of the existing files will be overwritten.
VIII Troubleshooting 6 USB USB 6 Symptom Data cannot be transferred normally. Corrective action • Make sure the USB 1.1 interface board and USB 1.1 hub used in the system are compliant with USB 1.1 standards, and that the USB 2.0 interface board and USB 2.0 hub used in the system are compliant with USB 2.0 standards. • Check that only one hub stage exists between the interface board and the DS-L1. Use of only one hub stage is recommended. • Make sure the USB cables used in the system are USB-compliant.
VIII Troubleshooting 6 • When connected to a PC, DS-L1 is recognized not as a storage unit but as some other USB • There may be a possibility that PictBridge-supporting printer is selected in Additional Settings screen in the SETUP menu. Select the selection mode for Mitsubishi DP900D printer, using "1.4.5 Settings for Direct Print" as a reference. device, such as a camera. Unable to print with a PictBridge-supporting printer.
VIII Troubleshooting 7 Microscope Control from the DS-L1 Microscope Control from the DS-L1 7 Symptom Corrective action (Or incorrect control • Use only the Nikon 80i/90i Microscope. This function cannot be used with a microscope from a different company or other Nikon microscope models. results.) • Turn on the power switch for the microscope. Unable to control microscope functions. • Wait until the microscope is ready for operation. It takes about 30 seconds for the microscope to be ready.
IX Technical Specifications 1 DS Camera Control Unit DS-L1 IX Technical Specifications 1 DS Camera Control Unit DS-L1 Display LCD display 6.
IX Technical Specifications 2 DS Camera Head DS-5M Recorded image Recorded image file format JPEG compliant, BMP 24-bit Image file size 2560 × 1920, 1280 × 960, 640 × 480 JPEG compression mode Fine, Normal, Basic, Draft Interface CF card Type I, Type II USB1.1 host interface USB mouse, USB keyboard, USB printer (CP900D) connectable, Nikon 80i/90i microscope USB2.
IX Technical Specifications 3 DS Cooled Camera Head DS-5Mc 3 DS Cooled Camera Head DS-5Mc Image pickup device 2/3-inch square, 5,240,000-pixel IT-CCD Frame rate 3.75 fps (5Mi), 7.5 fps (1.3Mp), 6.8 fps (1.3Mi), 15 fps (1.3Mp C.
IX Technical Specifications 6 Overall Specifications Storage environment Temperature –20° to 60° C Humidity 90%RH max. (No dewdrops allowed.) • • The contents of this manual are subject to change without notice. Due to an unavoidable characteristic of the CCD, there may be some missing pixels. Some of them may be automatically corrected, and the image may appear as if there are no missing pixels. Conforming standards This product conforms to the UL60950 standard.