User Manual

1. GENERAL
Operator's Manual WEE-1000 1.17
Name Functions
1. Antenna Diversity antenna
2. PHOTO UNIT connector Connects to the optional LS-901A Photo Control Unit.
3. ISOLATOR connector Connects to the telemetry unit by way of the SC-101A Isolator for direct wired
connection. DC power is supplied to the telemetry unit through the isolator.
4. DV UNIT connector Connects to the optional DV-101A Digital Video Unit through wired LAN.
5. PC connector Connects to the electroencephalograph through wired LAN.
6. POWER lamp Lights while AC power is supplied to the access point.
7. TX lamp Lights while in contact with the telemetry unit.
8. RX lamp Lights while receiving the data from the telemetry unit.
9. Protective ground terminal Use this terminal when protective grounding is required.
10. Equipotential ground Connects this terminal to the equipotential ground terminal on the wall with the
terminal ground lead when the equipotential grounding is required to ensure electrical
safety.
11. AC SOURCE socket Connects the power cord to supply AC power to the access point. When AC power
is supplied, the access point is turns on and the POWER lamp lights.
ZR-101AA Access Point
WARNING
Connect only the specified instruments to the connectors or socket marked with , by following the
specified procedure. Otherwise, electrical leakage current may harm the patient and operator.
Left side panel
Front panel
Right side panel
1
2
3
4
5
678
9
11
10