Owner's manual

NEXT>>>ADVANCE www.nextadvance.com SP500 series OEM Programmable Syringe Pump
NE500 series manual 37 Revision 101101
10.3 RS-232 Pump Network Connector Wiring
RS-232
To Network
To Computer
PC Com Port Connectors
25 Pin
9 Pin
3 - Receive 2 - Receive Connect to pump TXD
2 - Transmit 3 - Transmit Connect to pump RCV
9 - Ground 5 - Ground Connect to pump GND
10.4 Accessories
10.4.1 RS-232 Network Cables
RS-232 Network Primary Cable
7 foot cable, part #: CBL-PC-PUMP-7
25 foot cable, part #: CBL-PC-PUMP-25
Cable to connect a pump, or the first pump in a pump network, to a standard personal computer’s serial
port with a DB-9 or DB-25 connector. Included with this cable is a 9 pin to 25 pin converter.
RS-232 Network Secondary Cable
7 foot cable, part #: CBL-NET-7
25 foot cable, part #: CBL-NET-25
Cable to connect additional pumps, after the first pump, to the pump network.
10.4.2 Automation Cable: Reciprocating Pumps Control Cable
Part #: CBL-TTL-1
Using two SP-1000 family syringe pumps, this cable creates an automated continuous operation pumping
system, whereby one pump is refilling while the other is dispensing.
This cable is attached to two SP-1000 family syringe pumps via their ‘TTL I/O’ connectors. In this setup,
with the pumps configured for this operation, one pump will infuse while the other pump withdraws.
When one pump changes direction the other pump changes direction. When one pump starts, the other
pump starts. When one pump stops, for any reason, the other pump stops.
RCV from network
TXD to network
TXD to computer
RCV from computer