10.2 RS-232 Command Summary
<command> =>
DIA [ < float > ] Syringe inside diameter
PHN [ < phase data > ] Program Phase number
FUN [ < phase function > ] Program Phase function
< phase function > =>
RAT Pumping rate. ‘RATE’
INC Increment rate. ‘INCR’
DEC Decrement rate. ‘DECR’
STP Stop pump. ‘STOP’
JMP <phase data> Jump to Program Phase. ‘JP:nn’
PRI Program Selection Input. 'Pr:In'
PRL <count data> Program Selection Label definition. 'Pr:nn'
LOP <count data> Loop to previous loop start ‘nn’ times. ‘LP:nn’
LPS Loop starting Phase. ‘LP:ST’
LPE Loop end Phase. ‘LP:EN’
PAS <number data> Pauses pumping for ‘nn’ seconds. ‘PS:nn’
PAS [n.n] Pauses pumping for 'n.n' seconds. 'PS:n.n'
IF <phase data> If Program input low, jump to Program Phase. ‘IF:nn’
EVN <phase data> Set event trigger. ‘EV:nn’
EVS <phase data> Set event square wave trigger. 'ES:nn'
EVR Event trigger reset. ‘EV:RS’
BEP Sound short beep. ‘BEEP’
OUT { 0 | 1 } Set programmable output pin. ‘OUT.n’
RAT [ <float> [ UM | MM | UH | MH ] ] Pumping rate
VOL [ <float> ] Volume to be Dispensed
DIR [ INF | WDR | REV ] Pumping direction
RUN [ <phase data> ] Starts the Pumping Program
STP Stop/pauses the Pumping Program
DIS Query volume dispensed
CLD { INF | WDR } Clear volume dispensed
SAF [ <n> [ <n> [ <n> ] ] ] Safe communications mode
AL [ <on-off> ] Alarm mode
PF [ <on-off> ] Power failure mode
TRG [ FT | LE | ST ] Operational trigger mode
DIN [ 0 | 1] Directional input control mode
ROM [0 | 1] Pump Motor Operating TTL output mode
OUT 5 { 0 | 1 } Set TTL output level
IN { 2 | 3 | 4 | 6 } Query TTL input level
BUZ [ 0 | { 1 [ < n > ] } ] Buzzer control
VER Query firmware version
*ADR [ <n> [<n>] ] Network address (system command, valid regardless of
current address)
*RESET Clears program memory and resets communication
parameters (system command, valid regardless of current
address)