NE500 series manual 16 Revision 101101
Phase Function Rate Volume Direction
3
INC 1.0 0.1 ml Infuse
Phase Function
4
LOP 50
Phase Function
5
LPS
Phase Function Rate Volume Direction
6
DEC 1.0 0.1 ml Infuse
Phase Function
7
L0P 99
Phase Function Rate Volume Direction
8
DEC 1.0 0.1 ml Infuse
Phase Function
9
LPS
Phase Function Rate Volume Direction
10
INC 1.0 0.1 ml Infuse
Phase Function
11
LOP 50
Phase Function
12
JMP 02
7.4.4 Example 4: Complex Dispenses with External Synchronization
0
0
750
W
i
t
h
d
r
a
w
(
m
l
/
h
r
)
I
n
f
u
s
e
Start
Trigger
Start
Trigger
:60
:60
:60
Start
Trigger
Start over
Syringe
Refill
A more complex dispensing example, this Program contains different pumping requirements, including
dispenses with multiple pumping rates. The first set of 3 dispenses drops down to a lower pumping rate
during the dispense. When each dispense is completed, the buzzer beeps to alert the operator, then the
pump waits for a start trigger before starting the next dispense.
The next set of 3 dispenses have a fixed time interval of 60 seconds between dispenses. After the last set
of dispenses, the syringe is refilled by the amount infused, 17.25 ml. Then the buzzer beeps, to alert the
operator to the start of the first set of dispenses. The process is then repeated.
Phase Function Rate Volume Direction
1
RAT 750.0 ml/hr 0.5 ml Infuse
Phase Function Rate Volume Direction
2
RAT 300.0 ml/hr 1.5 ml Infuse