Manual

NEXT>>>ADVANCE www.nextadvance.com SP300 Programmable Syringe Pump
Publication #1200-01 11 03/23/10
9.5 Syringe Diameters and Rate Limits
Next Advance
www.nextadvance.com
Model: SP300 syringe pump
Maximum and minimum flow rates
Syringe
Manufacturer
Syringe
(cc)
Inside
Diameter
(mm)
Maximum
Rate
(mL/hr)
Minimum
Rate
(µl/hr)
Maximum
Rate
(mL/min)
B-D
1
4.699
39.27
0.73
0.6545
3
8.585
131
2.434
2.184
5
11.99
255.6
4.748
4.261
10
14.43
370.3
6.876
6.172
20
19.05
645.4
11.99
10.75
30
21.59
829
15.4
13.81
60
26.59
23.35
20
.95
HSW
1
39.11
0.727
0.6519
Norm-Ject
3
165.6
3.076
5
12.45
275.6
5.119
4.594
10
449.6
8.349
7.493
20
20.05
714.9
13.28
11.91
30
932.6
17.32
15.54
50
28.16
25.27
Monoject
1
58.59
1.088
0.9766
3
8.941
142.1
2.64
2.369
6
286.8
5.326
12
15.72
439.4
8.161
7.324
20
20.12
719.9
13.37
11.99
35
23.52
983.8
18.27
16.39
60
26.64
23.44
21.03
140 38.00 47.69 42.80
Terumo
1
4.7
39.28
0.73
0.6547
3
142.4
2.646
2.374
5
13
300.5
5.581
5.009
10
443.9
8.244
7.399
20
20.15
722.1
13.41
12.03
30
949
17.63
15.81
60
28.16
25.27
Air-Tite
10
449.6
8.349
7.493
20
20.25
729.2
13.55
12.15
30
900.3
16.72
15
50
29
27.78
24.92
Syringe
(µL)
Inside
Diameter
(mm)
Maximum
Rate
(µl/hr)
Minimum
Rate
(µl/hr)
Hamilton
0.5
0.103
18.86
0.001
Microliter
1
0.146
37.91
0.001
2
0.206
75.47
0.002
5
0.326
189
0.004
SGE
0.5
0.1
17.78
0.001
1
40.01
0.001