Instruction Manual

NEXT>>>ADVANCE www.nextadvance.com SP2200Programmable Syringe
Pump
Publication #1200-02 41 04/29/10
10. Appendix
10.1 RS-232 Command Summary
<command> =>
DIA [ < float > ] Syringe inside diameter
PHN [ < phase data > ] Program Phase number
FUN [ < phase function > ] Program Phase function
< phase function > =>
RAT Pumping rate.
INC Increment rate.
DEC Decrement rate.
STP Stop pump.
PAS <number data> Pauses pumping for ‘nn’ seconds, if 0, wait for start trigger.
PAS [n.n] Pauses pumping for 'n.n' seconds.
JMP <phase data> Jump to Program Phase.
PRI Sub-Program Selection Input from expansion connector
PRL <number data> Sub-Program Start Label definition.
LOP <count data> Loop to previous loop start ‘nn’ times.
LPS Loop starting Phase.
LPE Loop end Phase.
IF <phase data> If Program input low, jump to Program Phase.
EVN <phase data> Set event trigger.
EVS <phase data> Set event square wave trigger.
EPL <expansion pin> Pause program, wait for low edge or level on expansion pin
EPE <expansion pin> Pause program, wait for falling edge on expansion pin
EVE <expansion pin> Re-direct event trigger to expansion pin
EVR Event trigger reset.
OUT { 0 | 1 } Set programmable output pin.
OE0 <expansion pin> Set expansion pin low
OE1 <expansion pin> Set expansion pin high
TRG <n> Override Operational Trigger default setting
BEP Sound short beep.
RAT {C | I ] [ <float> [ UM | MM | UH | MH ]] Pumping rate
VOL [ <float> | <volume units>] Volume to be Dispensed, or set Volume units
DIR [ INF | WDR | REV ] Pumping direction
RUN [ <phase data> ] Starts the Pumping Program
[E [<phase data>] ] Pumping Program event trigger
STP Stop/pauses the Pumping Program
DIS Query volume dispensed
CLD { INF | WDR } Clear volume dispensed
SAF [ <n> [ <n> [ <n> ] ] ] Safe communications mode
AL [ <on-off> ] Alarm mode
PF [ <on-off> ] Power failure mode
TRG [ FT | FH | F2 | LE | ST | T2 | SP | P2] Operational trigger default mode
DIN [ 0 | 1] Directional input control mode
ROM [0 | 1] Pump Motor Operating TTL output mode
LOC [ <on-off> ] Lockout mode for Sub-Program Auto-Run
OUT [ E <expansion pin> | 5] { 0 | 1 } Set TTL output level
IN { E <expansion pin> | 2 | 3 | 4 | 6 } Query TTL input level
BUZ [ 0 | { 1 [ < n > ] } ] Buzzer control
VER Query firmware version
*ADR [ <n> [<n>] [ B <baud-rate>] ] Network address and baud rate (system command, valid
regardless of current address)