Instruction Manual

NEXT>>>ADVANCE www.nextadvance.com SP2200Programmable Syringe
Pump
Publication #1200-02 38 04/29/10
DB-9 Connector TTL I/O
Pin # Definition
Type Function
1 Vcc (5V) Reference Logic high reference. Power on indicator.
2 Operational Trigger
Command: TRG [aa]
Input
Configurable start/stop operational trigger input.
FT Foot Switch Falling edge: Start or stop trigger
FH Foot Switch Hold Falling edge: Start trigger
Rising edge: Stop trigger
F2 Foot Switch Reverse Rising edge: Start or stop trigger
LE Level Falling edge: Stop trigger
Rising edge: Start trigger
ST Start only Falling edge: Start trigger
T2 Start only Reverse Rising edge: Start trigger
SP Stop only Falling edge: Stop trigger
P2 Stop only Reverse Rising edge: Stop trigger
3 Pumping Direction Input Changes pumping direction according to setup
Command: DIN 0 DIN 1
Falling edge: Infuse Withdraw
Rising edge: Withdraw Infuse
4 Event Trigger Input Event input or user definable input
5 Program Output Output Program controlled output or user definable output
6 Program Input Input Program conditional input read by the “IF” program function.
Also user definable input.
Also used by the keypad lockout function.
7 Pump Motor Operating Output Setup Command: ROM 0
High: Pumping; Low: Not pumping
Setup Command: ROM 1
High: Pumping or Pause timer; Low: Not pumping
8 Pumping Direction Output High: Infuse; Low: Withdraw
9 Ground (0V) Reference Logic low reference
9.1 TTL I/O Operational Controls
Operational Trigger (Pin 2): The input signal on this pin controls the operation of the Pumping Program. Its
functionality is user configurable. Use the ‘TRG' command to configure the default settings for this input pin
(See Section 6.5, TTL I/O Operational Trigger Default Configuration).
Each option defines when the Operational Trigger input is activated. When activated, the trigger emulates a
‘Start/Stop’ key.
Foot Switch: Operates like a ‘Start/Stop’ key, whereby each falling edge (contact to ground) either starts
or stops/pauses the Pumping Program.
Foot Switch Hold: Falling edge starts the Pumping Program and the rising edge stops the Pumping Program.
With a foot switch, the Pumping Program will run while the foot switch was held down.
Foot Switch Reversed: Operates like the ‘Start/Stop’ key, whereby each rising edge either starts or
stops/pauses the Pumping Program.
Level Control: Falling edge stops/pauses the Pumping Program, Rising edge starts the Pumping Program.