User guide

NEXT>>>ADVANCE www.nextadvance.com SP1600/1800 Programmable Syringe Pump
Publication #1200-02 26 9/15/04
8.4.3 Example 3: Ramping the Flow Rate
Continuously ramp up and down the pumping rate. Starting at 200 ml/hr, the pumping rate will increment
to 250 ml/hr in 1.0 ml/hr steps after every 0.1 ml has been dispensed. Then the pumping rate will
decrement to 150 ml/hr in 1.0 ml/hr steps after every 0.1 ml has been dispensed. Finally, the pumping rate
is incremented back to 200 ml/hr in 1.0 ml/hr steps after every 0.1 ml has been dispensed, then the process
is repeated.
Phase Function Rate Volume Direction
1
RATE 200 ml/hr 0.1 ml Infuse
Phase Function
2
LP:ST
Phase Function Rate Volume Direction
3
INCR 1.0 0.1 ml Infuse
Phase Function
4
LP:50
Phase Function
5
LP:ST
Phase Function Rate Volume Direction
6
DECR 1.0 0.1 ml Infuse
Phase Function
7
LP:99
Phase Function Rate Volume Direction
8
DECR 1.0 0.1 ml Infuse
Phase Function
9
LP:ST
Phase Function Rate Volume Direction
10
INCR 1.0 0.1 ml Infuse
Phase Function
11
LP:50
Phase Function
12
JP:02