Pin # Definition Type Function
1 Vcc (5V) Reference Logic high reference. Power on indicator.
2 Operational Trigger Input
Configurable start/stop operational trigger input.
[Ft] Foot Switch Falling edge: Start or stop trigger
[FH] Foot Switch Hold Falling edge: Start trigger
Rising edge: Stop trigger
[F2] Foot Switch Reverse Rising edge: Start or stop trigger
[LE] Level Falling edge: Stop trigger
Rising edge: Start trigger
[St] Start only Falling edge: Start trigger
[t2] Start only Reverse Rising edge: Start trigger
[SP] Stop only Falling edge: Stop trigger
[P2] Stop only Reverse Rising edge: Stop trigger
[rL] Start on low level Low level: Start trigger
[rH] Start on high level High level: Start trigger
[SL] Stop on low level Low level: Stop trigger
[SH] Stop on high level High level: Stop trigger
[Et] Program function: redirect trigger to Event trap
3 Pumping Direction Input Changes pumping direction according to setup
[dr:rE] [dr:dU]
Falling edge: Infuse Withdraw
Rising edge: Withdraw Infuse
4 Event Trigger Input Event input or user definable input
5 Program Output Output Program controlled output or user definable output
6 Program Input Input Program conditional input read by the “IF” program function.
Also user definable input.
Also used by the keypad lockout function.
7 Pump Motor Operating Output [RUN.0] High: Pumping; Low: Not pumping
[RUN.1] High: Pumping or Pause timer
Low: Pumping Programmed stopped or paused
8 Pumping Direction Output High: Infuse; Low: Withdraw
9 Ground (0V) Reference Logic low reference
11.1 TTL I/O Operational Controls
While the user is changing settings or configuration from the keypad, external control by the ‘Pumping
Direction’ and ‘Operational Trigger’ inputs will be ignored. These controls will also be ignored if an alarm
condition exists.
Operational Trigger (Pin 2): The input signal on this pin controls the operation of the Pumping Program.
Its functionality is user configurable. Use the ‘TR:nn’ Setup Default Configuration to configure this input
pin (See Section 8.4, TTL I/O Operational Trigger Default Configuration).
Each option defines when the Operational Trigger input is activated. When activated, the trigger emulates
the ‘Start/Stop’ key:
Foot Switch: Operates like the ‘Start/Stop’ key, whereby each falling edge (contact to ground) either
starts or stops/pauses the Pumping Program.
Foot Switch Hold: Falling edge starts the Pumping Program and the rising edge stops the Pumping
Program. With a foot switch, the Pumping Program will run while the foot switch was
held down.