User Manual
NEXT>>>ADVANCE www.nextadvance.com SP1000 series operator’s manual
Manual SP1000 series 39
10.4.1 Program Function Commands
The following commands are relevant to the currently select Program Phase. Note: During a Pumping
Program’s operation, the currently selected Phase can change automatically.
PHASE NUMBER
PHN [ <phase data> ]
Set/query currently selected Program Phase:
Set:
Currently selected Phase is set to <phase data>. Previous Phase is stored in non-volatile memory
and the requested Phase is recalled from the non-volatile memory. Set is only valid if the
Pumping Program is not operating.
Query response:
<phase data> Currently selected Phase.
PUMPING PROGRAM FUNCTION
FUN [ <phase function> ]
Set/query the Pumping Program Phase’s function.
This command is relevant to the currently selected Phase. Set is only valid if the Pumping
Program is not operating
For a more detailed description of Program commands, see sec. 9.3, Program Function
Descriptions.
<phase function> =>
Rate Data Functions
When a Phase’s function is set to a “Rate Data Function”, use the ‘RAT’, ‘VOL’, and
‘DIR’ commands to setup the pumping parameters.
RAT Pumping rate. ‘RATE’
FIL Fill syringe to dispensed volume. ‘FILL’
INC Increment rate. ‘INCR’
DEC Decrement rate. ‘DECR’
Non-Rate Control Functions
STP Stop pump. ‘STOP’
PAS <number data> Pauses pumping for ‘nn’ seconds. ‘PS:nn’
PAS <n.n> Pauses pumping for 'n.n' seconds. 'PS:n.n'
PRI Sub-Program Selection Input. 'Pr:In'
PRL <number data> Sub-Program Start Label definition. 'Pr:nn'
LPS Loop starting Phase. ‘LP:ST’
LPE Loop end Phase. ‘LP:EN’
LOP <count data> Loop to previous loop start ‘nn’ times. ‘LP:nn’
JMP <phase data> Jump to Program Phase. ‘JP:nn’
IF <phase data> If Program input TTL pin low, jump to Phase. ‘IF:nn’
EVN <phase data> Set event trigger trap. ‘EV:nn’
EVS <phase data> Set event square wave trigger trap. 'ES:nn'
EVR Event trigger reset. ‘EV:RS’
TRG <nn> Override Operational Trigger Default Mode, set to <nn>,
‘tr:aa’
BEP Sound short beep. ‘BEEP’
OUT <TTL level> Set programmable output pin. ‘OUT.n’
PUMPING RATE
RAT [ C | I ] [ <float> [ <rate units> ] ]
Set/query pumping rate.