User Manual

NEXT>>>ADVANCE www.nextadvance.com SP1000 series operator’s manual
Manual SP1000 series 32
The pump then withdraws 0.25 ml (PH:07), pauses for 1 second (PH:08), then repeats this process if the
Program Input TTL line is low (PH:09), otherwise it continues with the next Phase.
Next, the pump pauses for 10 seconds (PH:10). Then it pauses again for the lesser of another 10 seconds
(PH:12) or until an Event Trigger occurs (PH:11). The Program then restarts (PH:13).
Phase Function
1
ET:RS
Phase Function
2
OUT.1
Phase Function Rate Volume Direction
3
RATE 800.0 ml/hr 5.0 ml Infuse
Phase Function
4
OUT.0
Phase Function
5
ET:07
Phase Function Rate Volume Direction
6
RATE 800.0 ml/hr 0.0 ml (OFF) Infuse
Phase Function Rate Volume Direction
7
RATE 1000.0 ml/hr 0.25 ml Withdraw
Phase Function
8
PS:01
Phase Function
9
IF:07
Phase Function
10
PS:10
Phase Function
11
ET:01
Phase Function
12
PS:10
Phase Function
13
JP:01
9.4.9 Example 9: Automatic Refill
While a foot switch is held, dispense continuously, starting with a 1.5 mL bolus. When the foot switch is
released, the syringe will refill to the volume that was dispensed.
In the pump’s general setup, set the TTL Trigger to “Foot Switch Hold” mode (TR:FH).
Phase Function
1
TR:ET Redirect trigger input to the Event Trap
Phase Function
2
ET:05 Set Trigger Event Trap to start syringe refill
Phase Function Rate Volume Direction
3
RATE 1000.0 ml/hr 1.5 mL Infuse Begin bolus dispense
Phase Function Rate Volume Direction
4
RATE 500.0 ml/hr 0.0 ml (OFF) Infuse Then infuse continuously
Phase Function Rate
5
FILL 1000.0 ml/hr Refill syringe when foot switch is released