User Manual
NEXT>>>ADVANCE www.nextadvance.com SP1000 series operator’s manual
Manual SP1000 series 30
Pumping Program. It is assumed that the sensor is positioned to refill the syringe with 60 ml. A withdraw
volume of 61 ml is set as a safety feature.
After the syringe is refilled, one 5 ml dispense is made every 5 hours. After 12 dispenses, the syringe is
refilled using the sensor again.
Phase Function
1
EV:03
Phase Function Rate Volume Direction
2
RATE 1000.0 ml/hr 61 ml Withdraw
Phase Function
3
LP:ST
Phase Function Rate Volume Direction
4
RATE 200.0 ml/hr 5.0 ml Infuse
Phase Function
5
LP:ST
Phase Function
6
LP:ST
Phase Function
7
PS:60
Phase Function
8
LP:60
Phase Function
9
LP:05
Phase Function
10
LP:12
Phase Function
11
JP:01
9.4.7 Example 7: Sub-Programs
This example shows some of the flexibilities provided by the Program Selection functions. The Pumping
Program starts by refilling the syringe with 50 ml at a fast pumping rate (Phase 1), then the Pumping
Program pauses for user sub-program selection (Phase 3). Then performs 5 dispenses of 10 ml at the
selected rates, then refills the syringe again and pauses for the next user sub-program selection.
The user is given the option of choosing one of three defined sub-programs.
1: Dispense 10 ml at 100 ml/hr (Phase 4)
2: Dispense 10 ml at 500 ml/hr (Phase 7)
3: Dispense 10 ml at 750 ml/hr (Phase 10)
After selecting the sub-program and pressing 'Start', the Pumping program continues execution at the
selected sub-program. After the 10 ml dispense, each sub-program jumps or continues with the loop
counter function (Phase 12). The first 4 loops continue Program Execution with the next user sub-program
selection. After the 5th loop, the program continues with Phase 13, which jumps back to the syringe refill
function and starts the whole program over.
Phase Function Rate Volume Direction
1
RATE 1500.0 ml/hr 50 ml Withdraw
Phase Function
2
LP:ST
Phase Function
3
PR:IN