User Manual
NEXT>>>ADVANCE www.nextadvance.com SP1000 series operator’s manual
Manual SP1000 series 26
Phase Function
3
LP:ST
Phase Function
4
LP:ST
Phase Function
5
PS:90
Phase Function
6
LP:03
Phase Function
7
BEEP
Phase Function
8
PS:30
Phase Function Rate Volume Direction
9
RATE 750 ml/hr 2.25 ml Infuse
Phase Function Rate Volume Direction
10
RATE 750 ml/hr 0.25 ml Withdraw
Phase Function
11
LP:EN
9.4.3 Example 3: Ramping the Flow Rate
m
l
/
h
r
R
a
t
e
Time
0
150
200
250
Phases
2 - 4
Phases
5 - 8
Phases
9-11
Note: Graph is representative only.
Pumping rate increments and decrements in steps.
Continuously ramp up and down the pumping rate. Starting at 200 ml/hr, the pumping rate will increment
to 250 ml/hr in 1.0 ml/hr steps after every 0.1 ml has been dispensed. Then the pumping rate will
decrement to 150 ml/hr in 1.0 ml/hr steps after every 0.1 ml has been dispensed. Finally, the pumping rate
is incremented back to 200 ml/hr in 1.0 ml/hr steps after every 0.1 ml has been dispensed, then the process
is repeated.
Phase Function Rate Volume Direction
1
RATE 200 ml/hr 0.1 ml Infuse
Phase Function
2
LP:ST
Phase Function Rate Volume Direction
3
INCR 1.0 0.1 ml Infuse
Phase Function
4
LP:50
Phase Function
5
LP:ST