User Manual
NEXT>>>ADVANCE www.nextadvance.com SP1000 series operator’s manual
Manual SP1000 series 25
9.3.18 'OUt.n': Set TTL Output Pin
n
Set ‘Program Output’ TTL I/O output (pin 5) to level ‘n’. If ‘n’ = 0, the output pin will be set
low. If ‘n’ = 1, the output pin will be set high.
9.3.19 'bEEP': Beep
Sounds a short beep.
9.4 Pumping Program Examples
9.4.1 Example 1: 2 Step Rate
ml Dispensed
0 5
30
0
2.5
500
ml/hr Rate
Infuse 5.0 ml at 500 ml/hr, and then infuse 25.0 ml at 2.5 ml/hr. Then stop the pump.
Phase Function Rate Volume Direction
1
RATE 500 ml/hr 5.0 ml Infuse
Phase Function Rate Volume Direction
2
RATE 2.5 ml/hr 25.0 ml Infuse
Phase Function
3
STOP
9.4.2 Example 2: Repeated Dispenses with Suck Back
m
l
/
h
r
R
a
t
e
Time
0
0
750
5:00
:30
Dispense 2.0 ml with a 5 minute pause between dispenses. In addition, after each dispense, a volume of
0.25 ml is sucked back to prevent dripping. Also, 30 seconds before the end of the pause interval, a beep is
sounded to alert the operator to prepare for the next dispense.
Starting with the second dispense, 0.25 is added to the volume dispensed to compensate for the sucked back
volume of the previous dispense. By changing the last Phase to a [LP:nn] function, the total number of
dispenses can be set.
When entering a function with associated data, such as with the ‘Pause’ in Phase 5, or the ‘Loop’ in Phase
6, the function is entered in 2 steps. First select the function and store it. Then enter the associated data.
Phase Function Rate Volume Direction
1
RATE 750 ml/hr 2.0 ml Infuse
Phase Function Rate Volume Direction
2
RATE 750 ml/hr 0.25 ml Withdraw