User Manual

NEXT>>>ADVANCE www.nextadvance.com SP1000 series operator’s manual
Manual SP1000 series 17
Setting
RS-232 Command Setting
FUN TRG <n>
<n> =>
Name Function
Ft 0 Foot Switch Falling edge starts or stops the Pumping Program
FH 1 Foot Switch Hold Falling edge starts the Pumping Program
Rising edge stops the Pumping Program
F2 2 Foot Switch Reversed Rising edge starts or stops the Pumping Program
LE 3 Level Control Falling edge stops the Pumping Program
Rising edge starts the Pumping Program
St 4 Start Only Falling edge starts the Pumping Program
t2 5 Start Only Reversed Rising edge starts the Pumping Program
SP 6 Stop Only Falling edge stops the Pumping Program
P2 7 Stop Only Reversed Rising edge stops the Pumping Program
rL 8 Start on Low Level Start Pumping Program on a low level
rH 9 Start on High Level Start Pumping Program on a high level
SL 10 Stop on Low Level Stop Pumping Program on a low level
SH 11 Stop on High Level Stop Pumping Program on a high level
Et 12 Trigger Event Pumping Program Function Only. Triggers an
event instead of stopping the pump.
8.5 TTL I/O Directional Control Input Configuration
aa
Configures how the TTL input ‘Pumping Direction’ (pin 3) will control the pumping
direction. (See sec. 11.1,
TTL I/O Operational Controls). The 2 letter configuration parameter to the right
of the colon (:) is defined as follows:
Setting Name Function
rE Reciprocating Pumps Falling edge: Infuse; Rising edge: Withdraw
dU Dual Pump Falling edge: Withdraw; Rising edge: Infuse
The setting names are relevant to a 2 pump system, whereby the ‘Directional Control Input’ TTL pin is
attached to the second pump’s ‘Pumping Direction Output’ TTL pin.
8.6 Pump Motor Operating TTL Output Configuration
n
Configures the functionality of the 'Pump Motor Operating' TTL output pin (TTL pin 7).
Settings: 0: Sets the output to logic high only when the motor is operating (pumping).
Sets the output to logic low when the motor is not operating or when the Pumping Program is
executing a pause timer or is stopped
1: Sets the output to logic high when the motor is operating (pumping) or when the Pumping
Program is executing a pause timer.
Set the output to logic low when the Pumping Program is stopped
8.7 Keypad Lockout
n
Setting: '0' = Disabled, '1' = Enabled.
*** The "Lockout Disable Key" needs to be inserted into the TTL I/O connector to display this setting***