User's Manual

BewellConnect - BW-CX10 - User Manual - 112014 BewellConnect - BW-CX10 - User Manual - 112014
34 35
EMC data
Guidance and manufacturer’s declaration - electromagnetic immunity - for equip-
ment and system that are not life-supporting. The model BW-CX10 is intended for
use in the electromagnetic environment specied below. The customer or the user of
the model BW-CX10 should assure that it is used in such an environment.
Immunity
test
EN 60601
test level
Compliance
level
Electromagnetic
environment - guidance
Conducted
RF EN
61000-4-6
3 Vrms
150 kHz
to 80
MHz
N/A
Portable and mobile RF communications equipment should be
used no closer to any part of the Infared Thermometer type BW-
CX10, including cables, than the recommended separation dis
-
tance calculated from the equation applicable to the frequency of
the transmitter. Recommended separation distance
d =
[
3.5
V1
P
]
150 MHz to 80 MHz
d =
[
3.5
E1
P
]
80 MHz to 800 MHz
d =
[
7
E1
P
]
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from xed RF transmitters, as determined by an
electromagnetic site survey, should be less than the compliance
level in each frequency range. b Interference may occur in the vi
-
cinity of equipment marked with the following symbol:
Radiated
RF EN
61000-4-3
3 V/m
80 MHz
to 2.5
GHz
3 V/m
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
aected by absorption and reection from structures, objects and people.
a. Field strengths from xed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to xed RF transmitters, an electromagnetic site survey should be consid-
ered. If the measured eld strength in the location in which the Infared Thermometer type
BW-CX10 is used exceeds the applicable RF compliance level above, the Infared Thermome-
ter type BW-CX10 should be observed to verify normal operation. If abnormal performance
is observed, additional measures may be necessary, such as reorienting or relocating the
Infared Thermometer type BW-CX10.
b. Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 3V/m.
Recommended separation distances between portable and mobile RF communications
equipment and the equipment or system - for equipement and systems that are not
life-supporting.
The model BW-CX10 is intended for use in an electromagnetic environ-
ment in which radiated RF disturbances are controlled. The customer or the user of the
model BW-CX10 can help prevent electromagnetic interference by maintaining a mini-
mum distance between portable and mobile RF communications equipment (transmitters)
and the model BW-CX10 as recommended below, acording to the maximum output power
of the communications equipment.
Rated maximum output
of transmitter
w
Separation distance according to
frequency of transmitter m
150 MHz to 80 MHz
[
3.5
V1
P
]
d =
80 MHz to 800 MHz
[
3.5
E1
P
]
d =
800 MHz to 2.5 GHz
[
7
V1
P
]
d =
0.01 - 0.12 0.23
0.1 - 0.38 0.73
1 - 1.2 2.3
10 - 3.8 7.3
100 - 12 23
For transmitters rates at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmit-
ter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800MHz, the reparation distance for the higher frequency range
applies.
NOTE 2 these guidelines may not apply in all situations. Electromagnetic propagation is
aected by absorption and reection from structures, objects and people.
BW-CX10_MANUAL_FR-UK_072015.indd 34-35 13/07/2015 17:11