EMC data
Guidance and manufacturer’s declaration - electromagnetic immunity - for equip-
ment and system that are not life-supporting. The model BW-CX10 is intended for
use in the electromagnetic environment specied below. The customer or the user of
the model BW-CX10 should assure that it is used in such an environment.
Immunity
test
EN 60601
test level
Compliance
level
Electromagnetic
environment - guidance
Conducted
RF EN
61000-4-6
3 Vrms
150 kHz
to 80
MHz
N/A
Portable and mobile RF communications equipment should be
used no closer to any part of the Infared Thermometer type BW-
CX10, including cables, than the recommended separation dis
-
tance calculated from the equation applicable to the frequency of
the transmitter. Recommended separation distance
d =
[
3.5
V1
P
]
√
150 MHz to 80 MHz
d =
[
3.5
E1
P
]
√
80 MHz to 800 MHz
d =
[
7
E1
P
]
√
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from xed RF transmitters, as determined by an
electromagnetic site survey, should be less than the compliance
level in each frequency range. b Interference may occur in the vi
-
cinity of equipment marked with the following symbol:
Radiated
RF EN
61000-4-3
3 V/m
80 MHz
to 2.5
GHz
3 V/m
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
aected by absorption and reection from structures, objects and people.
a. Field strengths from xed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to xed RF transmitters, an electromagnetic site survey should be consid-
ered. If the measured eld strength in the location in which the Infared Thermometer type
BW-CX10 is used exceeds the applicable RF compliance level above, the Infared Thermome-
ter type BW-CX10 should be observed to verify normal operation. If abnormal performance
is observed, additional measures may be necessary, such as reorienting or relocating the
Infared Thermometer type BW-CX10.
b. Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 3V/m.
Recommended separation distances between portable and mobile RF communications
equipment and the equipment or system - for equipement and systems that are not
life-supporting.
The model BW-CX10 is intended for use in an electromagnetic environ-
ment in which radiated RF disturbances are controlled. The customer or the user of the
model BW-CX10 can help prevent electromagnetic interference by maintaining a mini-
mum distance between portable and mobile RF communications equipment (transmitters)
and the model BW-CX10 as recommended below, acording to the maximum output power
of the communications equipment.
Rated maximum output
of transmitter
w
Separation distance according to
frequency of transmitter m
150 MHz to 80 MHz
[
3.5
V1
P
]
√
d =
80 MHz to 800 MHz
[
3.5
E1
P
]
√
d =
800 MHz to 2.5 GHz
[
7
V1
P
]
√
d =
0.01 - 0.12 0.23
0.1 - 0.38 0.73
1 - 1.2 2.3
10 - 3.8 7.3
100 - 12 23
For transmitters rates at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmit-
ter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800MHz, the reparation distance for the higher frequency range
applies.
NOTE 2 these guidelines may not apply in all situations. Electromagnetic propagation is
aected by absorption and reection from structures, objects and people.