10. TECHNICAL CHARACTERISTICS & PRECISION
Product Type Infra-red medical thermometer
Model MyThermo BW-CX10
Brand BewellConnect
Classication Classe IIa
Temperature sensor IR Sensor
Wavelenght 5-14µm
Emissivity 0.95
Normal condition of use Room temperature
Body mode :15°C à 40°C
Surface mode : 15°C à 40°C
Normal condition of storage Température : From -20°C to 50°C
Humidity : 10 % RH ~ 85%RH
Atmospheric pressure 101.325Kpa
Humidity rate 10 %RH ~ 85 %RH
Battery DC3.7V (Lithium battery)
Size 60.7 x 37.4 x 96.3 mm
Weight 117 g
Mesurement display °C / °F
Measuring range Body mode : From 30° to 43°C (86.0°F to 109.4°F)
Surface mode : From 0° to 93.3°C (32°F to 199.9°F)
Room mode : From 0° to 50°C (32°F to 122°F)
Repeatability ≤±0.3°C
Measuring distance 5 to 8 cm
Display resolution 0.1°C
Accuracy +/- 0.2°C in Body mode (35°C-42°C)
+/- 1.0°C in Room mode
+/- 0.4°C in Surface mode
Bluetooth Technology 4.0
Consumption < 50 mW
Voltage 3.6V~4.2V
Current in sleep mode ≤15A
Current when turned on ≤150mA
Automatic stop 10~20 secondes
LONGEVITY USE: The MyThermo BW-CX10 is guaranteed for 40.000 readings.
11. MYTHERMO BW-CX10 ACCURACY
From 34°C à 35.9°C = +/-0.3°C
ISO 80601-2-56:2009From 36°C à 39°C = +/-0.2°C
From 39.1°C à 42.5°C = +/-0.3°C
MyThermo BW-CX10 can take temperature measurement below 30°C or above 43°C.
Outside of these mesurement ranges, accurancy cannot be guaranteed
12. SUPPLIED ACCESSORIES
- User Manual
- Quickstep
- Charging cable
- Base
BW-CX10_MANUAL_FR-UK_072015.indd 28-29 13/07/2015 17:11