Product Manual
Page 7 of 21
Draft 1.1
• Wrap appropriate cuffs at each site. Attach the hoses from the Cuff Link Control
Box to each cuff as shown above. The green connectors go to the arm cuffs and
yellow to the ankles.
Brachial Pressure: Use the computer cursor and click on the site you would like to
begin with, usually the right brachial.
Place the Doppler probe at an angle to the skin over the radial artery. *HINT* Hold
the probe close to the end. It also helps to rest your hand on the patient to keep the
probe in place. One of the keys to a successful exam is being able to keep the probe in
place as you inflate and deflate. If it moves you will not be able to hear the Doppler
sounds return and you will have to repeat the inflation. Use the angled vascular adapter
(Cheaterâ„¢) to get the best signal and to keep the probe from moving during inflation
and deflation. Move the probe slowly back and forth until the strongest arterial signal is
heard.