User guide
New Era Pump Systems Inc.
www.SyringePump.com
Model NE-500/NE-501 Multi-Phaser™
Publication #1200-02 24 09/18/14
Phase Function
2
LPS
Phase Function
3
PRI
Phase Function
4
PRL 01
Phase Function Rate Volume Direction
5
RAT 100.0 mL/hr 10 mL Infuse
Phase Function
6
JMP 12
Phase Function
7
PRL 02
Phase Function Rate Volume Direction
8
RAT 500.0 mL/hr 10 mL Infuse
Phase Function
9
JMP 12
Phase Function
10
PRL 03
Phase Function Rate Volume Direction
11
RAT 750.0 mL/hr 10 mL Infuse
Phase Function
12
LOP 05
Phase Function
13
JMP 01
7.4.8 Example 8: Dispensing with Complex Synchronization
:01
Repeat if
Program Input
low
:10
:10
Restart on Event Trigger
or 10 seconds
Event Trigger
W
i
t
h
d
r
a
w
I
n
f
u
s
e
TTL Program Ouput
TTL High TTL Low
Event Trigger
Infuse until
Event Trigger
This example demonstrates a complex interaction with external equipment, such as synchronizing with another
syringe pump. The Program includes a variety of interactions with external equipment, which demonstrates the
various control possibilities of the NE-500/501.
The Pumping Program begins by canceling any previous event traps (PHASE 01) and raising the ‘Program