Model NE-500/NE-501 Multi-Phaser™
Publication #1200-02 21 09/18/14
Phase Function
5
LOP 02
Phase Function Rate Volume Direction
6
RAT 750.0 mL/hr 0.5 mL Infuse
Phase Function Rate Volume Direction
7
RAT 300.0 mL/hr 1.5 mL Infuse
Phase Function
8
BEP
Phase Function
9
LPS
Phase Function
10
PAS 60
Phase Function Rate Volume Direction
11
RAT 500.0 mL/hr 3.75 mL Infuse
Phase Function
12
LOP 03
Phase Function Rate Volume Direction
13
RAT 900.0 mL/hr 17.25 mL Withdraw
Phase Function
14
BEP
Phase Function
15
PAS 00
Phase Function
16
LPE
7.4.5 Example 5: Control from a High-Low Pressure Sensor
10
25
m
l
/
h
r
Pumping
Rate
Program
Output Pin
Event Input
Pin
Pressure high sensor
Pressure low sensor
Pressure trigger points
This example demonstrates a Pumping Program whose control depends on an external sensor. Assuming a
pressure sensor that is configured to detect a high pressure point and a low pressure point, the Pumping
Program individually selects whether it will react to the high or low pressure point.
The “Program Output” pin on the TTL I/O connector (pin 5) is used to select the high or low pressure point.
When low, the low pressure point is selected (PHASE 01), and when high, the high pressure point is selected
(PHASE 05). The Program begins by infusing continuously at 10.0 mL/hr (PHASE 02), while a background
trap is set for the low pressure point (PHASE 03). To create a delay when the pressure sensor is switched from
high pressure to low pressure when the “Program Output” pin is set, a small volume is pumped (PHASE 02,
06) before the background traps are set.