Owner manual
New Era Pump Systems Inc.
www.SyringePump.com
Model NE-4000 Multi-Phaser™
Publication #1200-01
37
09/05/14
VOLUME DISPENSED
DIS
Queries volume dispensed only. Set not applicable.
Response:
I <float> W <float> <volume units>
Where: “I <float>“ refers to the infusion volume dispensed, and “W <float>” refers to the withdrawn
volume.
CLEAR VOLUME DISPENSED
CLD { INF | WDR }
Sets the Infused or withdrawn volume dispensed to 0. Command is only valid while the Pumping
Program is not operating. Query is not applicable.
INF = Infusion volume
WDR = Withdrawn volume
Query is not applicable.
9.4.3 Configuration and Setup Commands
New settings for any of the following commands will be stored in the non-volatile memory.
LOW MOTOR NOISE
LN [ <on-off> ]
Set/query low motor noise mode. Increases motor stepping resolution.
ALARM SETUP
AL [ <on-off> ]
Set/query alarm setup mode. Set alarm enables or disables alarm buzzer mode.
POWER FAIL SETUP
PF [ <on-off> ]
Set/query Power Failure mode. Set Power Failure enables or disables Power Failure mode.
TTL I/O OPERATIONAL TRIGGER DEFAULT SETUP
TRG [ <trigger setup> ]
Set/Query TTL I/O Operational Trigger (Pin 2) default input configuration.
TTL I/O Operational Trigger is set to <trigger setup>.
<trigger setup> => FT = Foot switch trigger (falling edge start/stop)
FH = Foot switch hold (falling edge start, rising edge start)
F2 = Foot switch reverse (rising edge start/stop)
LE = Level trigger (rising edge start, falling edge stop)
ST = Start only trigger (falling edge start)
T2 = Start only trigger reversed (rising edge start)
SP = Stop only trigger (falling edge stop)
P2 = Stop only trigger reversed (rising edge stop)
RL = Start on low level
RH = Start on high level
SL = Stop on low level
SH = Stop on high level
OF = Trigger off (disabled)