Manual
New Era Pump Systems Inc.
WWW.SYRINGEPUMP.COM
Model NE-1600/NE-1800 Multi-Phaser™
Publication #1200-02
43 8/1/14
Pin # Definition Type Function
1 Vcc (5V) Reference Logic high reference. Power on indicator.
2 Operational Trigger Input
Configurable start/stop operational trigger input.
[Ft] Foot Switch Falling edge: Start or stop trigger
[FH] Foot Switch Hold Falling edge: Start trigger
Rising edge: Stop trigger
[F2] Foot Switch Reverse Rising edge: Start or stop trigger
[LE] Level Falling edge: Stop trigger
Rising edge: Start trigger
[St] Start only Falling edge: Start trigger
[t2] Start only Reverse Rising edge: Start trigger
[SP] Stop only Falling edge: Stop trigger
[P2] Stop only Reverse Rising edge: Stop trigger
[rL] Start on low level Low level: Start trigger
[rH] Start on high level High level: Start trigger
[SL] Stop on low level Low level: Stop trigger
[SH] Stop on high level High level: Stop trigger
[OF] Trigger off (disabled)
[Et] Program function: Redirects trigger to Event trap
[bt] Program function: Redirects ‘Stop’ key to Event trap
3 Pumping Direction Input Changes pumping direction according to setup
[dr:rE] [dr:dU]
Falling edge: Infuse Withdraw
Rising edge: Withdraw Infuse
4 Event Trigger Input Event input or user definable input
5 Program Output Output Program controlled output or user definable output
6 Program Input Input Program conditional input read by the “IF” program function.
Also user definable input.
Also used by the keypad lockout function.
7 Pump Motor Operating Output [RUN.0] High: Pumping; Low: Not pumping
[RUN.1] High: Pumping or Pause timer
Low: Not pumping
8 Pumping Direction Output High: Infuse; Low: Withdraw
9 Ground (0V) Reference Logic low reference
10.1 TTL I/O Operational Controls
While the user is changing settings or configuration from the keypad, external control by the ‘Pumping
Direction’ and ‘Operational Trigger’ inputs will be ignored. These controls will also be ignored if an alarm
condition exists.
Operational Trigger (Pin 2):
The input signal on this pin controls the operation of the Pumping Program. Its
functionality is user configurable. Use the ‘TR:nn’ Setup Default Configuration to configure this input pin.
(See Section 7.4, TTL I/O Operational Trigger Default Configuration)
Each option defines when the Operational Trigger input is activated. When activated, the trigger emulates the
‘Start/Stop’ key:
Foot Switch:
Operates like the ‘Start/Stop’ key, whereby each
falling
edge (contact to ground) either starts
or stops/pauses the Pumping Program.
Foot Switch Hold:
Falling
edge starts the Pumping Program and the
rising
edge stops the Pumping Program.
With a foot switch, the Pumping Program will run while the foot switch was held down.
Foot Switch Reversed:
Operates like the ‘Start/Stop’ key, whereby each
rising
edge either starts or
stops/pauses the Pumping Program.