Manual
New Era Pump Systems Inc.
WWW.SYRINGEPUMP.COM
Model NE-1600/NE-1800 Multi-Phaser™
Publication #1200-02
26 8/1/14
8.4.4 Example 4: Complex Dispenses with External Synchronization
0
0
750
W
i
t
h
d
r
a
w
(
m
l
/
h
r
)
I
n
f
u
s
e
Start
Trigger
Start
Trigger
:60
:60
:60
Start
Trigger
Start over
Syringe
Refill
A more complex dispensing example, this Program contains different pumping requirements, including
dispenses with multiple pumping rates. The first set of 3 dispenses drops down to a lower pumping rate during
the dispense. When each dispense is completed, the buzzer beeps to alert the operator, then the pump waits for
a start trigger before starting the next dispense.
The next set of 3 dispenses have a fixed time interval of 60 seconds between dispenses. After the last set of
dispenses, the syringe is refilled by the amount infused, 17.25 mL. Then the buzzer beeps, to alert the operator
to the start of the first set of dispenses. The process is then repeated.
Phase Function Rate Volume Direction
1
RATE 750.0 mL/hr 0.5 mL Infuse
Phase Function Rate Volume Direction
2
RATE 300.0 mL/hr 1.5 mL Infuse
Phase Function
3
BEEP
Phase Function
4
PS:00
Phase Function
5
LP:02
Phase Function Rate Volume Direction
6
RATE 750.0 mL/hr 0.5 mL Infuse
Phase Function Rate Volume Direction
7
RATE 300.0 mL/hr 1.5 mL Infuse
Phase Function
8
BEEP
Phase Function
9
LP:ST
Phase Function
10
PS:60
Phase Function Rate Volume Direction
11
RATE 500.0 mL/hr 3.75 mL Infuse
Phase Function
12
LP:03
Phase Function Rate Volume Direction
13
RATE 900.0 mL/hr 17.25 mL Withdraw
Phase Function
14
BEEP