Manual
New Era Pump Systems Inc. WWW.SYRINGEPUMP.COM Model NE-1600/NE-1800 Multi-Phaser™
Publication #1200-02 i 8/1/14
Quick Start Instructions
Assumes that the pump was not previous programmed with a multiple Phase Pumping Program.
• Plug in the pump.
• Press the power switch to turn on power.
• Press any key to stop the display from blinking.
Setup Pumping Parameters
To Change Numbers:
• Use the up-arrow keys to increment individual digits.
•
To set/clear the decimal point
: Simultaneously press the 2 up-arrow keys under the 2 digits next to the decimal
point position. Alternatively, press and hold the left-most up-arrow key for at least 1 second. When the digit
increments from 9 to 0, the decimal point will begin to shift. Release the key when the decimal point is correct.
• Press any non-arrow key, or wait 2 seconds, to enter the new setting. The display will blink when a new value is
entered and stored in memory.
Set the Syringe Inside Diameter:
• Momentarily press the ‘Diameter’ key. Set the diameter.
Set the Pumping Rate.
• Momentarily press the ‘Rate’ key.
• To change the
pumping rate units
:
− Momentarily press the ‘Rate’ key again. The display will show:
− Press any up-arrow key to select the next available rate units while the units LEDs are blinking
− .Press any non-arrow key, or wait 2 seconds, to set the rate units.
• Set the pumping rate. If the pumping rate is out of range, the display will show:
Set the Volume to be Dispensed or Continuous Pumping
• Momentarily press the ‘Volume’ key.
• When the display shows: , the pump is set for continuous pumping. Pressing any up arrow key will change
the display to 0.
• For continuous pumping: Set the volume to 0.
• For a Volume to be Dispensed: Set the volume. The default units are set according to the syringe diameter.
Set the Pumping Direction
• When the ‘Withdraw’ LED is lit, the pump is set for withdrawing. When not lit, the pump is set for infusing. Use the
‘ ’ key to change the pumping direction.
Load the Syringes
• Press in the white drive-nut button to move the pusher bar.
• Insert the syringe plungers in the pusher bar bracket slots. Space unused syringe slots evenly around center.
• Insert the syringe barrel flange in the flange bracket slots with the syringe barrel holder lowered onto the syringes.
Tighten the flange bracket onto the syringe flanges.
Prime / Purge:
Press and hold the ‘Start/Stop’ key for one second. Release to stop.
Start the Pump:
Press and release the ‘Start/Stop’ key to start or stop the pump.
When Pumping
• The pumping rate can be changed.
• With continuous pumping, the pumping direction can be changed.
PUMP RESET:
PUMP RESET:PUMP RESET:
PUMP RESET:
Press and hold the right-most up-arrow key while turning on power to the pump.