User Manual

Table Of Contents
Using the Field Programmer
Notes:
8-44 N_SIGHT R900 User’s Manual for the Neptune Handheld System
After you select a
format, all the values
ass
ociated with it appear
as shown in Figure 8.40.
You can then make any
necessary changes.
Figure 8.40 Gas Values Shown on
New Format Screen
5 Touch or click in MULTIPLIER to select the correct value
from 0.01 to 1000 to be use for the format.
6 Type a value for INPUT to
enter the input rate. This is controlled
by the index.
7 Type the 9-digit number for the Pressure Con
figuration Factor
index in PCF.
Refer to the listing of PCF indexes, refer to “Pressure Configuration
Factor Indexes,” in Appendix C.These are examples only
. This list
does not include every index a utility has
in its system. If there are
any questions about the correct pressure configuration factor,
contact the meter manufacturer.
8 Touch in DispDig
its to select one of the following:
4 = Initial Reading will be in range of 0000 - 9999
5 = Initial Reading will be in range of 00000 - 99999
6 = Initial Reading will be in range of 000000 - 999999