User Manual

Table Of Contents
Using the Field Programmer
8-24 N_SIGHT R900 User’s Manual for the Neptune Handheld System
4 Do one of the following:
Touch or click Refresh to force the R900G to tran
smit the
most current reading.
Touch or click Sleep to place the R900G endpoint in sleep
mode to be used for shipping.
Touch or click Wakeup to activate the R900
G endpoint from
sleep mode.
•Click Close to exit the Program Gas screen and return to the
start screen.
Reading the R900G Endpoint Register
The Field Programmer is designed to be used as a programming and
troubleshooting tool rather than a reading device. However, the
Field Programmer is capable of collecting readings.
To
read a gas register, com-
plete the following steps.
1 Select GA
S from the
Nept
une Programmer
System start screen.
2 Select the Read Gas tab.
The Gas Read screen
appears. See Figure 8.26.
Figure 8.26 Gas Read Screen
Any time you are using the Field Programmer gas program you can use the
functions on the Command Gas screen.
Pilot Release