Preferences dialog - Display tab
Lock OSD (On Screen Display) controls when calibrated - selects if the monitor’s On Screen Display control buttons
are locked when the monitor is calibrated. Locking the controls can be useful to prevent accidental or unauthorized
adjustment of the monitor once it is calibrated. Options are:
Don’t lock
Lock all except Bright and Contrast
Lock all
Note: The controls will be unlocked if the calibration is turned o in the main window.
Use Auto Luminance (if supported) - enables the Auto Luminance sensor circuit inside the display monitor to increase
the Intensity (brightness) stability as the display ages and warms up. This feature will not be used for Targets with
“Maximum Possible” Intensity selected, or if the Target Intensity falls outside the range that the Auto Luminance circuit
is capable of controlling. This feature is only available on the following models: LCD2490WUXi, LCD2490WUXi2,
LCD2690WUXi, LCD2690WUXi2, and LCD3090WQXi.
The MultiSync PA Series of displays (except the PA231W) automatically always use Auto Luminance to maintain the
stability of the monitor. The PA231W does not support Auto Luminance.
See the “Features by Model” on page 57 for all model specic functionality.
Note: This setting will be disabled if no relevant displays are connected.