Specifications
English-29
English
STAND-ALONE CALIBRATION
Stand alone calibration can recalibrate the display without using a computer by the integrated front sensor or connecting an
external sensor. There are 4 functions in STAND-ALONE CALIBRATION.
• Self Calibration: Manually adjust the integrated front sensor with an external near range color analyzer reference, which is
traceable to a primary standards laboratory.
• Gamma Adjust: Recalibrates Gamma data by the integrated front sensor.
• DICOM Measurement: Measures DICOM response.
• AMBIENT SENSOR CALIBRATION: To manually adjust the integrated ambient light sensor with an external color analyzer
reference, which is traceable to a primary standards laboratory.
NOTE: Stand before proper Stand-Alone calibration can be performed, the display should warm-up for a minimum of
30 minutes. If Stand-Alone calibration is started before the monitor is warmed up, a warning will appear on-screen
(Figure A.1).
NOTE: Refer to the KEY MAP (Figure A.2) to adjust the settings for calibration.
Stand-Alone calibration should be performed in either portrait or landscape orientation.
NOTE: To start calibration without a computer, perform the following steps.
When the “NO SIGNAL” message appears on screen, press the “LEFT” and then “RIGHT” buttons and hold them
down simultaneously. The CALIBRATION menu will appear.
Figure A.2
KEY MAP
UP/DOWN: Changes from setting to setting
LEFT/RIGHT: Changes the setting selection
(ie...SELF or COPY)
SELECT: Moves to the next step of calibration
EXIT: Moves back one step of calibration
Figure A.1
CAUTION: Please use a properly-managed external near range color analyzer. If there is any doubt about the result of
the Stand-Alone calibration result, please contact your supplier.