Performing the DICOM conformance test
This function is available only when the display has been calibrated to DICOM GSDF.
Performing the DICOM conformance test ensures DICOM conformance of the display by measuring and checking the display
properties.
The DICOM conformance test is recommended to be performed after 60 minutes of continuous lighting time of the backlight of
the LCD panel. If the test is performed in less than 60 minutes of lighting time, a margin of measurement error may be larger
because luminance is not stable enough.
HINT: If the power management function of the display works in less than 60 minutes due to the computer power setting,
change the computer setting, or try “Advanced OSD Controls without video input”.
Procedures of DICOM conformance test:
1. Checking the gamma setting
The (+) / (-) buttons are located in the back of the display. Press either button twice, and the present operating state appears.
Make sure the gamma is set to “DICOM GSDF”.
Gamma setting
Present operating state
P R E S E N T S T A T E
L UMI NANCE : 5 0 1
AMB L IGHT : 300
lx
INPUT : Di spl ayPor t
CAL I BRAT I ON SETT I NG
CONF I G : 1
L UMI NANCE : 5 0 0
GAMMA : DI COM GSDF
COLOR : 6500K
2. Displaying the DICOM conformance test screen
Press either button again, and the DICOM conformance test screen appears.
Make sure at least 60 minutes of continuous lighting time of the backlight has elapsed.
DICOM conformance test
A. Result of the last test
( “---” indicates that the test has not been performed.)
B. Luminance status
Stable: Luminance is stable. (Recommended)
Unstable: Luminance is unstable.
C. Continuous lighting time of the backlight
Over 60 minutes (Recommended)
“Over 240 min” indicates more than 240 minutes has elapsed.
3. Performing the DICOM conformance test
With the present operating state displayed, press and hold both the buttons for 2 seconds.
The screen that shows the test is in execution appears.
Screen that shows the test is in execution
Completing the measurement takes between 30 seconds to 1 minute.
To cancel the measurement, press and hold the both plus (+) and minus (-) buttons for over 2 seconds.