Item Brochure

Features
EasyOne Pro LAB offers improved efficiency and quality
of care by providing complete pulmonary function
testing on site
Multiple-Breath N
2
washout for FRC, RV, TLC, LCI
Single Breath CO Diffusion (DLCO)
Spirometry
Flexible EMR capabilities
No warm-up time
High-resolution color touch screen
Automatic and fast calibration for each test performed
Automated ATS/ERS interpretation
Challenge testing with manitol and methacholine protocols
Wide range of data export options
Easy raw data export for scientific use
No service contracts
Easy software upgrades
Remote support capability
Specifications
Flow/Volume Sensor
Type: Ultrasonic transit-time
R
ange: ±16 l/s
Resolution: 4 ml/s
Accuracy: ±2% or 0.02 l/s
Resistance: < 1.5 cmH2O/l/s at 12 l/s
CO Sensor
Type: Non-dispersive infrared
Range: 0 to 0.35%
Resolution: 0.0001%
Accuracy: < 0.001%
Tracer Gas Sensor (Helium, N
2
)
Type: Ultrasonic transit-time
Range: 0 to 50%, 0 to 100%
Resolution: 0.01%, 0.05%
Accuracy: 0.05%, 0.2%
CO
2
Sensor
Type: Non-dispersive infrared
Range: 0 to 15%
Resolution: 0.005%
Accuracy: ±0.05% (from 0 to 5%)
Standards
Quality System: ISO 9001, ISO 13485
FDA: 510(k) market clearance
MDD 93/42/EEC: CE marked
Electrical: EN 60601-1, EN 60601-1-2
Electrical and Environmental Requirements
Supply: 100 to 240 VAC, 50/60 Hz
Temperature: 10 to 40°C
Humidity: 30 to 75% (non condensing)
Normal
D
L
,CO
>
L
L
N
DL
,CO
>
L
L
N
DL
,CO
>
L
L
N
Y
e
s
N
o
Y
e
s
N
o
Y
e
s
N
o
VC
>
L
L
N
VC
>
L
L
N
TLC
>
L
L
N
TLC
>
L
L
N
Yes
N
o
Normal Restriction Obstruction Mixed
defect
PV
disorders
CW and NM
disorders
ILD
Pneumonitis
Asthma
CB
Emphysema
N
o
N
o
N
o
N
o
Y
e
s
Y
e
s
Y
e
s
Y
e
s
FEV1/VC
>
L
L
N
ATS/ERS Interpretation Decision Tree
© Copyright 2013 ndd Medical Technologies, Inc. All rights reserved. rev. 03/14