Probe compensation
When you connect the probe to an input channel for the first time, you must adapt the probe to the input
channel. An uncompensated or incorrectly compensated probe will result in measurement errors. Perform
the probe compensation as follows:
1. Set the damping factor of the probe to 10X in the menu, also set the switch on the probe to 10X and
connect the probe to channel 1. When using the hook tip, make sure that it remains securely
connected to the probe. Connect the probe tip to the signal connector of the probe compensator and
connect the clamp of the reference cable to the earth terminal of the probe compensator; then press
the AUTOSET button.
2. Check the displayed waveforms and adjust the probe until correct compensation is achieved
Overcompensated Correct compensationUndercompensated
3. Repeat the process if necessary to obtain the most uniform image possible.