- 108 -
Notes on the battery
Battery status indicator
Loading the oscilloscope
Connect the mains cable to a power source. Switch on the unit with the power switch - on the back of the
unit (make sure the "-" side is pressed down). If the
battery status indicator on the control panel is yellow, the battery is charging. When fully charged, the indicator
lights up green. The lithium battery may not be fully charged when you receive the unit. Therefore, please
charge the battery for 12 hours before using it for the first time. The battery will last up to 4 hours after fully
charging, depending on use.
A battery indicator appears at the top of the screen when the oscilloscope is running on battery power and
shows the battery charge status.
The empty battery symbol indicates that the battery will only supply power for a maximum of 5 minutes.
Hint:
To avoid overheating the battery during charging, the ambient temperature must not exceed the value
specified in the technical data.
Replacing the lithium battery
Under normal conditions, it is not necessary to replace the battery. However, if this should become necessary,
the replacement can only be carried out by qualified personnel; a lithium battery with the same technical data
must be used.