User's Manual
Chapter 2 Function Reference — Config_Alarm_Deadband
NI-DAQ FRM for PC Compatibles 2-66
©
 National Instruments Corporation
Figure 2-2. 
Low Alarm Deadband
Config_Alarm_Deadband
 is a high-level function for NI-DAQ event messaging. Because 
this function uses the current inputRange and polarity settings to translate triglevel and 
deadbandWidth from volts to binary, you should not call 
AI_Configure and change these 
settings after you have called 
Config_Alarm_Deadband. For more information on NI-DAQ 
event messaging, see the low-level function 
Config_DAQ_Event_Message. When you are 
using this function, the analog input data acquisition must be run with interrupts only 
(programmed I/O mode). You cannot use DMA. See 
Set_DAQ_Device_Info for how to 
change modes.
trigLevel +
deadbandWidth/2
trigLevel -
deadbandWidth/2
trigLevel
On: low alarm on
Off: low alarm off
on
off
Time