User's Manual
Chapter 2 Function Reference — Calibrate_E_Series
NI-DAQ FRM for PC Compatibles 2-62
©
 National Instruments Corporation
Calibration Constant Loading by NI-DAQ
NI-DAQ automatically loads calibration constants into calDACs whenever you call functions 
that depend on them (
AI, AO, DAQ, SCAN, and WFM functions). The following conditions apply:
12-bit E Series Devices 16-bit E Series Devices
• The same set of constants is 
correct for both polarities of 
analog input.
• One set of constants is valid for 
unipolar, and another set is valid 
for bipolar configuration of the 
analog output channels. When you 
change the polarity of an analog 
output channel, NI-DAQ reloads 
the calibration constants for that 
channel.
• Calibration constants required by the 16-bit 
E Series devices for unipolar analog input 
channels are different from those for bipolar 
analog input channels. If you are acquiring 
data from one channel, or if all of the channels 
you are acquiring data from are configured for 
the same polarity, NI-DAQ selects the 
appropriate set of calibration constants for 
you. If you are scanning several channels, and 
you mix channels configured for unipolar and 
bipolar mode in your scan list, NI-DAQ loads 
the calibration constants appropriate for the 
polarity that analog input channel 0 is 
configured for.
• Analog output channels on the 
AT-MIO-16XE-50 can be configured only for 
bipolar operation. Therefore, NI-DAQ always 
uses the same constants for the analog output 
channels.